Emerson Daniel 3410 Series Operation Manual page 144

Gas ultrasonic flowmeters
Hide thumbs Also See for Daniel 3410 Series:
Table of Contents

Advertisement

Troubleshooting comm., mech., and elec.
January 2019
Table C-1: Mechanical/Electrical troubleshooting
Problem
Connected
communication line to
the flow computer but
no signal is received
Communicating with
meter but all chords
display failures
Chord is not indicating
144
Solution(s)
Check for loose connections at the flow meter and the flow computer
(see
3410 Series engineering
Check the CPU Module, Field Connection board and the Power Supply
wiring. Ensure the terminal block wiring and connectors are making
good contact.
Verify that the resistance of transducers is within Specification (1-2
Ω).
Check the Acquisition board.
Check the interconnect cables between the base enclosure and the
transmitter enclosure.
Check the resistance of the failed transducer.
If Chord A is not indicating, change the transducer cables from Chord
D to chord A
If Chord D then fails, the transducers are bad on Chord A.
The same test procedure can be accomplished by swapping Chords B
and C (or E and H, or F and G) if a chord fails in either chord.
Note
Note that the outside chord cables cannot be exchanged with inner chord
cables.
Operations manual
DAN-20057316
(continued)
drawings).
Operations manual