Operations manual
DAN-20057316
Figure 5-4: Calibration - Meter Factor: Polynomial
Three coefficients are specified for each flow direction from the Daniel MeterLink
Calibration > Meter Factors page as shown in the table below:
Table 5-4: Data points for Polynomial wet calibration
Daniel Meterlink
Display Name
Forward C0
Forward C1
Forward C2
Forward C3
Model 3418 GUSM
™
Data points, options and guidelines
Data points affected
•
FwdC
0 Options:
•
Enter a value (m/s or ft/s) within the range [-1, 1 m/s].
Data points affected
•
FwdC
1 Options:
•
Enter a value (dimensionless) within the range [0.95, 1.05].
Data points affected
•
FwdC
2 Options:
•
Enter a value (s/m or s/ft ) within the range [-0.1, 0.1 s/m].
Data points affected
•
FwdC
0 Options:
•
Enter a value (m/s or ft/s) within the range (s
the range [-0.1, 0.1 s
2
/m
2
2
/m
].
Configurations
January 2019
2
2
2
or s
/ft
) within
47