Emerson Daniel 3410 Series Operation Manual page 143

Gas ultrasonic flowmeters
Hide thumbs Also See for Daniel 3410 Series:
Table of Contents

Advertisement

Operations manual
DAN-20057316
electronic/electrical troubleshooting techniques, using lap top computers as well as digital
volt/ohm meters. Great care should be taken not to "short out" a given electronic/electrical
circuit when Troubleshooting.
Table C-1: Mechanical/Electrical troubleshooting
Problem
No power the unit
Cannot communicate
with Daniel MeterLink
program
One or more of the
chords is not indicating
a reading (reporting
zeros)
Waveform contains an
excessive amount of
noise
Model 3418 GUSM
Solution(s)
Check for correct voltage (AC or DC) to the input of the Field
Connection Board. (See
Wiring Diagram).
Check the main power source for blown fuse or tripped circuit
breaker. Reference your "as built" installation drawings for your
location.
Check the fuses on the Field Connection Board. Reference fuse F1 and
F2 locations.
Ensure that the meter is properly powered.
Ensure that the computer cable is properly connected to the field
connection board (see
connector and check your interface pins (RS-485 or RS-232).
Verify that the communication parameters of the Daniel MeterLink
program are set according to jumpers on the meter CPU board. See
Daniel MeterLink utilities
configuring communications.
Check switch S-1 on the field connection board and ensure that it is in
the correct position (RS-485 or RS-232 communication).
Note
Note that S1 is located at the top left side of the board.
Check for loose connections at the cable connectors. See
engineering
drawings.
Check the resistance of the transducers (should be approximately 1-2
Ω).
Problem also may be caused by a bad Acquisition board or
interconnect cable. See
information.
Check system status in Daniel MeterLink
flagged errors.
Check the CPU board LEDs.
Increase the StackSize until noise level decreases (settings can be 1
(none), 2, 4, 8, or 16). If increasing the StackSize is not successful, try
turning on the filter or consult with Flow Lifecycle Services for Daniel
products if you are unsure of how stacking a signal can effect the meter's
operation. Refer to Technical Support under the Help menu of Daniel
MeterLink for contact information.
Troubleshooting comm., mech., and elec.
3410 Series engineering
drawings, System
3410 Series engineering
drawings) for the
of this manual for instructions on
3410 Series engineering drawings
, Meter > Monitor for any
January 2019
3410 Series
for more
143