Error recovery for program download failure
If the Program Download failed during firmware upgrade, Daniel MeterLink can no longer
connect to the meter or a message Attempt FTP-only connection... displays when trying to
connect perform a Backup Upgrade.
It is possible for the meter to lose its programming if the meter power fails during a
firmware upgrade process. If this occurs, use the following special procedure, Backup
Upgrade, to connect to the meter and download the program again.
Backup upgrade
Procedure
1. In Daniel MeterLink
Model 3418 GUSM
upgrade is complete you will be able to reconnect to the meter with Daniel
MeterLink
.
prompts you with a message that it must disconnect from the
meter. Once the firmware upgrade is complete you will be able to reconnect to the
meter with Daniel MeterLink
Important
When the meter restarts, it takes about two minutes before you will be able to
reconnect depending on the firmware upgrade being performed. If the database
does need to be re-initialized, it could take up to five minutes.
program downloads. If all the program components are successfully updated, they
will show to be the same date and version as the Currently Installed Versions and
the Download button will be disabled. If one or more components are still not
updated, click Download to continue the upgrade process. Repeat the process to
ensure all components are updated.
CAUTION
POWER LOSS DURING PROGRAM DOWNLOAD
The meters program may be lost if the meter loses power in the middle of a
program download.
If this occurs perform the Backup Upgrade procedure to attempt a connection to
download the program again.
, go to the File pull-down menu and select Program Settings.
The Program Settings dialog displays. Enable the Allow FTP-only connection
checkbox and click OK to close Program Settings.
a. For Serial Port Connections: Ensure you are connected to Port A. You may
need to adjust your File > Meter Directory settings for the connection so that
they match the port default settings. Port A will default to 19200 baud with a
modbus address of 32.
b. For Ethernet Connections: If you are connecting to the meter over an
Ethernet port, you should be able to connect with the same IP address as
normal. If this is unsuccessful, the meter may have defaulted to
.