Configurations
January 2019
Note
Note that the meter provides two sets of dry calibration coefficients - one set for each flow
direction.
Dry calibration procedure
Prerequisites
NOTICE
Modification of dry calibration parameters is not recommended. The dry calibration
parameters should not be modified unless directed by Daniel personnel.
Procedure
1. Select the Calibration > Meter Factors menu.
2. Click the Advanced View button to display the zero flow calibration polynomial
3. Set the flow calibration method to None.
4. Enter the appropriate forward and reverse meter factors and click Write to send the
5. Click the Cancel button to cancel any changes and close the dialog box.
40
Figure 5-1: Calibration - Meter factors page
coefficients. Regardless of what flow calibration method is selected, the zero flow
calibration polynomial coefficients will be applied first.
factors to the meter.
The meter provides separate dry calibration coefficients for each flow direction as
listed in the table below.
Operations manual
DAN-20057316