Figure 7-5: Waveform viewer - frequency domain
Zero crossing and first motion markers
Two markers display along the horizontal axis for either the Stacked or Filtered waveform.
If filtering is on, the markers will be on the Filtered waveform. If filtering is off, the markers
will be on the Stacked waveform. The green marker shows the point where the first
motion is detected. The purple marker shows the zero crossing which is the point that the
meter uses as the arrival point of the signal.
Navigate the waveform viewer
Procedure
1. Select the waveforms to collect from the Meter box and click Read. Selections
124
include:
All Chords
Chord A
Chord B
Chord C
Chord D
Chord E
Operations manual