Commands
January 2019
7.1.6
Baseline viewer
Use the Baseline viewer to monitor the meter's flow direction and flow characteristics
including:
Flow Velocity
Profile Factor
Swirl Angle
Symmetry
Cross-flow
Path Turbulences
Access the Baseline viewer from the Meter > Monitor (Detailed) dialog while connected to
your meter.
Note
You must have a valid Continuous Flow Analysis key to enable this feature. See
optional keys
the baseline parameters for your meter using the Tools > Set Baseline Wizard menu.
Figure 7-8: Baseline viewer
The Baseline viewer indicates the current flow direction, forward or reverse, and gauges
display the flow characteristics of the meter. The Baseline viewer dialog displays a
horizontal "green zone" that indicates the expected value range. As an example,
depicts out-of-range values for turbulence on all four chords.
Read the baselines viewer gauge values
Table 7-2
Table 7-2: Baseline viewer gauge values
Value
Avg Velocity
130
for instructions about this topic and
describes the values for each gauge and how the green zones are defined.
Description
Displays the average flow velocity for the meter. This gauge contains no
green zone.
Operations manual
Set Baseline parameters
Operations manual
DAN-20057316
Obtaining
to configure
Figure 7-8