8. Click Save to save the configuration file. A default filename is provided based on the
9. Click Print to print the configuration currently opened in the editor. Only the
10. Use Convert to download the configuration from a legacy Daniel Ultrasonic meter
Model 3418 GUSM
Figure 7-3: Compare Configurations Options dialog
Exclude registers that are typically different: includes items that could change
regularly such as SpecFlowPressure, SpecFlowTemperature, Gas composition,
etc. A complete list is defined in reg_list_compare_config_exclude.txt found
under the installation directory for Daniel MeterLink.
Include read-only registers: includes, in the compare, read-only registers stored
in the collected configurations. This includes points like firmware version, K-
factors, etc.
meter name, time, and date the configuration was collected. You can keep the
default name or change it. The file is saved, by default, to the Data folder directory
defined under Program Settings. You may change the directory location if desired.
registers in the currently selected view will be printed. You can also click the print
icon,
, on the tool bar to print the opened configuration.
to later generation of electronics.
a) First connect to the legacy electronics. Collect and save the configuration
using the Edit/Compare Configuration dialog box. Upgrade the electronics
on the meter.
b) Connect to the new meter electronics and open the Edit/Compare
Configuration dialog. Open the configuration collected from the legacy
electronics. The Convert button is enabled.
c) Click Convert to read the configuration from the new meter and modify it
with the data from the legacy configuration. Daniel MeterLink displays this
modified configuration in the dialog. At this point, Daniel MeterLink has not
written anything to the new meter. All the values highlighted in yellow are
values from the legacy configuration.
d) Click Write Checked to write the changed portion of the configuration to the
new meter. If any data point failed to be written, you must correct the invalid
value and click Write Checked again until the configuration is written
without any errors. Once the configuration is written, you can choose to
compare the displayed configuration with the configuration in the meter by
selecting Meter and clicking Compare. This will verify everything was written
correctly.