Installing the ACCQPrep AI 5
AutoInjector
Overview
This instruction sheet will guide you through the
installation of the ACCQPrep AI 5 AutoInjector into the
ACCQPrep HP125 system.
Tools Required:
1/4" wrench
●
#2 Phillips screwdriver
●
#1 Phillips screwdriver
●
Side cutters
●
Parts List:
Part
Part Number
AutoInjector
60-5237-098
Panhead Screws
231514303
Lock Washers
233210400
72" Black Tubing
023050309
15" Teflon Tubing
023050204
Cable tie
489010100
Installation
1. Turn O
power to the ACCQPrep HP125 system.
FF
2. Disconnect the power cord from the back of the
unit.
3. Remove the cover panel on the right side of the
instrument (Figure 1, L).
4. Carefully cut the plastic tie securing the AutoIn-
jector power supply cable to the back of the
panel (Figure 1, R).
Figure 1: Cover panel (L) and cable tie (R)
Instruction Sheet #69-5233-836
5. Connect the AutoInjector power cable to the
port on the back of the syringe drive and secure
with the provided panhead screws and internal
lock washers (Figure 2).
Note
If the power supply cable isn't long enough, you may
need to access the back of the AutoInjector through
the back panel of ACCQPrep HP125 system.
Figure 2: Location of screw placement on the
syringe drive
6. Once the power supply cable is secured, you
now slide the AutoInjector Module into
may
the ACCQPrep system. (Figure 3).
Figure 3: AutoInjector partially installed
7. Using the screws (removed in Step 3) attach the
AutoInjector assembly into place.
8. The back panel may be reinstalled on the system
if it was necessary to remove it.
March 2018