Figure 3-2. Sample Probe Assembly Installation.
Figure 3-2 illustrates the sample probe assembly installation. The instructions
below can be used for the standard probe or the uptake capillary of the C-Flow
PFA nebulizer.
Insert the probe through the cylindrical support collar.
1
Using just your fingers, tighten the thumbscrew on the side of the collar to
secure the probe in the collar.
2
Insert the probe into the Y-arm assembly.
3
NOTE
The Y-arm assembly has two holes at either end. Use the smaller hole if you are
using the C-Flow PFA nebulizer; reverse the Y-arm assembly to use the larger
hole if you are using the standard probe.
NOTE
The support collar simply rests on top of the end of the Y-arm assembly. If the
end of the probe hits the edge of a sample vial or a solid part of the rinse
station block, the probe and collar will be pushed upward without bending (and
possibly damaging) the probe.
Loosen the thumbscrew on the collar and set the probe a little higher than
where you think it needs to be. Later on, after the probe height has been
4
checked, you can lower the probe.