Load the autosampler sample tray with an empty sample rack.
1
If multiple sample vial racks (different numbers of positions) will be used,
select the sample rack containing the maximum number of positions for the
initial verification run.
For information about placing the sample rack, see Chapter 3, "Installing the
Autosampler."
Turn the autosampler power switch on and verify that the LED power
indicator is on.
2
The LED power indicator is green in color. The indicator will flash upon power
up and will then remain lit when the unit is at the ready state. It is located on
the front of the unit at the top right corner of the base, directly above the
power switch.
Using the ICP-MS software, designate the sample positions at the left rear, left
front, right rear, and right front of the sample tray.
3
Figure 4-1. Sample Test Positions.
Place sample vials at the four corners.
4
Command the autosampler to move the sample probe to the designated
sample positions. Check that the sample probe correctly accesses each position
5
and that the probe descends into the center of each sample vial.
NOTE:
If the autosampler alignment is not correct, contact Teledyne CETAC
Technologies Customer Service and Support or an authorized representative.
ASX-112FR Autosamplers Operator's Manual
Chapter 4: Verifying Installation
45