ASX-112FR Autosampler Operator's Manual
Chapter 8: Operating a CETAC Autosampler Using a Terminal Program
Figure 8-11
Turn on the autosampler.
11
The HyperTerminal window should display an OK.

Autosampler Commands

The following commands will produce various responses from the
autosampler.
Command
Ver
Home
Tray=n
Tube=row-col-down
Pmp on
Pmp off
Rinse
Down=n
Up
Table 2 Autosampler Commands
You can use these commands to determine if the CETAC autosampler is
communicating and functioning properly. If more assistance is needed, please
contact CETAC customer service.
76
"ASCII Setup" Window
Description
Returns firmware version.
Returns all axes to home position, same
as power up.
Defines tray size and n= #of positions.
Tube=row-column-down as defined by
tray command. Example: Tube=3-4-
150
Pump on if unit has a pump.
Pump off if unit has a pump.
Moves sample probe to the rinse
position, extends and retracts the probe
3 times and starts rinse pump. Stays in
down position with pump running.
up,pmp off stops the pump.
Moves the Z-axis down by the
parameter(n) in mm. Do not run Down
command if probe is not all the way up
on up position or damage may occur to
probe or Z-axis.
Moves Z-axis to upper most postion.