Thorough Cleaning - Teledyne CETAC ASX-112FR Operator's Manual

Autosampler
Table of Contents

Advertisement

ASX-112FR Autosampler Operator's Manual
Chapter 6: Maintaining the Autosampler
Shut down and unplug the autosampler power cord.
1
For information about shutting down the autosampler, see page 50.
Remove the cover.
2
Remove the sample and standards vials and the sample rack.
3
Wipe the rotary tray, spill tray, autosampler cabinet, autosampler arm, and
cover using a damp clean-room wiper.
4
Wipe again with a new clean-room wiper.
5
This process removes any remaining contaminants.
Dry the rotary tray, spill tray, autosampler cabinet, autosampler arm, and cover
using a new clean-room wiper.
6
Avoid scratching the cover while drying. The autosampler must be thoroughly
dry before you turn the power on.
Although external cleaning removes spills and contaminants from most of the

Thorough Cleaning

autosampler components, it may be necessary to clean the autosampler more
thoroughly. It may also be necessary to perform a more thorough cleaning if a
large spill occurs. To do so, complete the following steps:
Shut down and unplug the autosampler power cord.
1
Remove the cover.
2
Remove the sample and standards vials and the sample rack.
3
Remove the rotary tray.
4
For information about removing the sample tray, see "Replacing the Rotary
Tray" on page 57.
Do not allow liquids to run or drip inside the tray seal when the rotary tray is
removed.
Remove the rinse station.
5
For information about removing the rinse station, see "Replacing the Rinse
Station" on page 58.
Wipe loose particles from the top of the autosampler cabinet and exposed
parts of the mechanism with a dry, lint-free cloth.
6
Clean the entire cabinet top surface area under the cover.
Use a damp clean-room wiper to wipe the interior and exterior of the cover,
the autosampler cabinet and spill tray, and all autosampler components until
7
they are clean.
56