Integra CUSA Clarity Instructions For Use Manual page 4

Bone tip pack
Hide thumbs Also See for CUSA Clarity:
Table of Contents

Advertisement

EN – ENGLISH
2. Place the handpiece on the torque base as shown
and make sure that the 2 flat metal sides fit snugly
in the metal slot at the end of the base.
3. Hold the handpiece in the base and place the
provided torque wrench (color side towards
handpiece) over the tip.
CAUTION
To avoid product damage, always use the torque base
to hold the handpiece while using the torque wrench to
tighten or loosen the tip. Avoid over torqueing the tip.
4. Rotate the Torque wrench clockwise until it clicks
twice.
NOTE:
Keep the torque wrench and base in the sterile field in
case you need to change a tip during the procedure.
Attaching the Nosecone and the Flue
1. Slide the larger end of the nosecone over the tip and
slide it down over the neck of the handpiece making
sure the white dots are aligned.
2. Rotate the nosecone clockwise until it locks into
place and the white dot on the nosecone aligns
with the white dot on the body of the handpiece.
Make sure the white dots are aligned on the nosecone
and handpiece.
3. Slide the flue onto the nosecone base. Make sure the
flue does not cover the distal end, abrasive surface, of
the tip.
Priming the System
Once you have completed system setup, prime the
irrigation system as described in the CUSA Clarity
Ultrasonic Surgical Aspirator System Operator's
Manual. Repeat Prime until you see irrigation
fluid at the handpiece tip.
Testing the Handpiece
Once you have completed priming the system, test the
handpiece to ensure that you have assembled the tip
correctly. Refer to the CUSA Clarity Ultrasonic Surgical
Aspirator System Operator's Manual for instructions
on how to test the handpiece.
CAUTION
Federal (US) law restricts this device to sale by or on the
order of a licensed healthcare practitioner.
Sterility is guarenteed unless the package is opened
or damaged.
4