Starting Reinfusion - Fresenius Medical Care multiFiltrate Instructions For Use Manual

Table of Contents

Advertisement

4.10.7 End of treatment
4.10.7.1 Terminating the treatment

4.10.7.2 Starting reinfusion

Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
 Press the [ESC] key.
 Use the rotary selector to select End of
treatment from the menu bar and press
[OK] to confirm.
The blood pump is running.
 Stop the treatment by pressing the
[STOP] key.
Press the [STOP] key for approx.
3 seconds.
The venous clamp closes.
The blood pump is stopped.
or
 Use the [ESC] key to select a different
menu from the menu bar.
 Continue the treatment by pressing the
[START/RESET] key.
or
 Disconnect the arterial patient line from
the patient and connect it to the NaCl
solution.
 Use the rotary selector to select Start
disconnection? [OK] to confirm! and
press [OK].
When selecting Disconnect without blood
return, the program will go directly to
"Disconnecting the patient" without reinfusion
after a safety prompt (see chapter 4.10.8 on
page 4-103).
Chapter 4: Operation
Haemoperfusion (HP)
4-101