Chapter 12: Specifications
Voltage dips, short
interruptions, and
voltage variations
according to
IEC 61000-4-11
Power frequency
magnetic fields
(50 / 60 Hz)
according to
IEC 61000-4-8
Note: U
is the AC voltage before the test level is applied.
T
Conducted RF
disturbances
according to
IEC 61000-4-6
Radiated RF
disturbances
according to
IEC 61000-4-3
12-8
< 5% U
(> 95% dip
T
of the
) for
U
T
0.5 periods
40% U
(60% dip of
T
the
) for 5 periods
U
T
70% U
(30% dip of
T
the
) for 25 periods
U
T
< 5% U
(> 95% dip
T
of the
) for
U
T
5 seconds
3 A/m
3 V
eff
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
< 5% U
(> 95% dip
T
of the
) for
U
T
0.5 periods
40% U
(60% dip of
T
the
) for 5 periods
U
T
70% U
(30% dip of
T
the
) for 25 periods
U
T
< 5% U
(> 95% dip
T
of the
) for
U
T
5 seconds
3 A/m
3 V
eff
3 V/m
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
The quality of the supply voltage should
correspond to that of a typical commercial
and / or hospital environment. If the
operator of the multiFiltrate device requires
continuous operation, even in the event of
interruptions in the energy supply, it is
advisable to supply the multiFiltrate using
an uninterruptible power supply or a
battery.
Power frequency magnetic fields should
correspond to the typical values found in
the commercial and hospital environment.
Portable and mobile radio devices should
not be any closer to the multiFiltrate device
(including the lines) than the
recommended safety distance, which is
calculated according to the relevant
equation for the transmission frequency.
Recommended safety distance:
d = 1.2 √P
For 150 kHz to < 80 MHz
d = 1.2 √P
For 80 MHz to < 800 MHz
d = 2.3 √P
For 800 MHz to 2.5 GHz
Where P is the nominal power of the
transmitter in watts (W) in accordance with
the information from the transmitter
manufacturer and d is the recommended
safety distance in meters (m).
In accordance with an investigation on
a
site
, the field strength of stationary
transmitters should be lower than the
compliance level for all frequencies
Interference is possible in the
vicinity of devices which have this
symbol.
b
.