Connecting the patient

Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Tip
If the patient is not yet available, then recirculation can be continued by
pressing the [START/RESET] key.
CRRT treatments with citrate anticoagulation
If using an NaCl solution with two
connectors:
 Disconnect the arterial patient line from
the NaCl solution and connect it to the
arterial vascular access.
or
 Connect the arterial and venous patient
line to the vascular access.
If using an NaCl solution with one
connector:
 Connect the arterial and venous patient
line to the vascular access.
and
 Use the rotary selector to select Start
connection? [OK] to confirm! and press
[OK].
The blood pump will deliver at the
programmed rate.
Settable rate:
10 ml/min to 100 ml/min
(default 100 ml/min)
The citrate pump will deliver at the
corresponding rate.
It will stop if the optical detector senses
opaque fluid or, for safety reasons, after
10 minutes at the latest.