CRRT treatments with citrate anticoagulation
 Disconnect the venous patient line from
the patient.
 Remove the calcium pump segment from
the stationary part (stator) of the calcium
pump at the pump segment adapter.
The rotor of the calcium pump
immediately starts to remove the pump
segment (2 revolutions).
 Remove the citrate pump segment from
the stationary part (stator) of the citrate
pump at the pump segment adapter.
The rotor of the citrate pump
immediately starts to remove the pump
segment (2 revolutions).
Support the removal of the pump
segment by slightly pulling on it.
If a Ci-Ca pump segment cannot be removed
completely, press the pump segment adapter
into the appropriate pump once again and
repeat the removal process.
 Remove the pump segment adapter from
the groove.
 Press and hold the [START/RESET] key
until the pump segment has been
completely removed.
Support the removal of the pump
segment by slightly pulling on it.
 Remove the heater bag (depending on the
treatment mode) from the heater.
Open the clamps before and after the
heater bag to facilitate removal of the
bag.
 Remove and dispose of the tubing
system.
To remove the remaining pump segments,
proceed as described above.
 Use the rotary selector to select View
treatment history? [OK] to confirm!
Confirm with [OK].