4.8.8.2

Terminating the treatment

4.8.8.3

Starting reinfusion

Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Membrane plasma separation (MPS)
 Stop the treatment by pressing the
[STOP] key.
Press the [STOP] key for approx.
3 seconds.
The venous clamp closes.
The blood pump is stopped.
 Continue the treatment by pressing the
[START/RESET] key.
or
 Disconnect the arterial patient line from
the patient and connect it to the NaCl
solution.
 Use the rotary selector to select Start
disconnection? [OK] to confirm! and
press [OK].
When selecting Disconnect without blood
return, the program will go directly to
"Disconnecting the patient" without reinfusion
after a safety prompt (see chapter 4.8.9 on
page 4-79).
Chapter 4: Operation
4-77