2.12.5 Warnings relating to consumables and accessories

Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
The alarm limits of the patient leakage current for type CF applied parts
are as follows:
– 10 µA AC / DC (normal case, error-free)
– 50 µA AC / DC (the first time an error occurs)
Please address any queries to the local service support organisation.
Warning
Consumables must be discarded after the treatment in compliance with
the regulations for the disposal of potentially contaminated materials.