Complete Tubing Arrangement; Preparation; Priming The Tubing System - Fresenius Medical Care multiFiltrate Instructions For Use Manual

Table of Contents

Advertisement

4.9.5

Preparation

4.9.5.1

Priming the tubing system

Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Note
The patient must not be connected during preparation. If the optical
detector senses opaque fluid during the preparation phase, the system
asks whether a patient has been connected.
Slow continuous ultrafiltration (SCUF)
This screen will be displayed immediately if
the setting in the System parameters menu
was set to Complete tubing arrangement.
 Confirm the completely inserted tubing
system using the rotary selector to select
Set up? [OK] to confirm! and press
[OK].
 Use the rotary selector to select Start
priming? [OK] to confirm! and press
[OK].
Heparin can be added to the NaCl solution
used for priming and rinsing via the Bolus
menu field.
Chapter 4: Operation
4-83

Advertisement

Table of Contents
loading

Table of Contents