4.10.5.3 Recirculation / waiting for patient
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Note
The filter life, which is shown in the status bar, is automatically added to
the service life of the tubing systems.
After the rinse is completed and the
pre-defined volume used, the blood pump will
stop.
Audible signal
If using an NaCl solution with two
connectors:
 Disconnect the venous patient line from
the empty bag and connect it to the NaCl
solution.
If using an NaCl solution with one
connector:
 Connect the venous patient line to the
arterial patient line using the recirculator.
 Continue recirculation by pressing the
[START/RESET] key.
Chapter 4: Operation
Haemoperfusion (HP)
4-97