Chapter 4: Operation

Treatment menu

4.11 Treatment menu

4.11.1 Deselecting Ci-Ca anticoagulation

4-104
Warning
After deselecting citrate anticoagulation, the operator has to ensure an
alternative anticoagulation procedure.
Warning
After deselecting citrate anticoagulation, the CVVHD treatment may
only be continued / performed if a calcium-containing HF solution /
dialysate is used.
Warning
After deselecting citrate anticoagulation, the Ci-Ca lines must not be
removed from the pumps before the patient has been completely
disconnected.
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
 Press the [ESC] key.
 Use the rotary selector to select
Treatment from the menu bar and press
[OK].
The menu fields shown depend on the
selected treatment mode.
 Use the rotary selector to select the
required parameter field from the menu
field and press [OK].
After a timeout, the display will automatically
return to the treatment main screen.