Alarm priorities
Audible alarm
suppression /
mute
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Follow-up alarms of the same priority are not signaled by the system.
Follow-up alarms of higher priority are signaled in cases where the
system has not already reached a safe condition.
Silencing the audible alarm will always suppress an alarm for a duration
of 2 minutes. Any new alarm that occurs will activate the silenced
audible alarm.
Chapter 5: Alarm processing
5-5