Chapter 7: Functional description
Legend
Treatment parameters
Pre-Post CVVH
Blood flow
Ultrafiltration
UF goal
Cont. heparin adm.
Sub predilution
Sub postdilution
Temperature
7-4
1
Blood pump
2
Haemofilter
3
Air detector
4
Venous clamp
5
Filtrate pump
6
Blood leak detector
7
Filtrate
8
Scales 3 and 4
9
Predilution
10
Postdilution
11
Check valve
12
Heater
13
Substituate pump
14
Scales 1 and 2
15
Substituate
Min.
10
Off / 10
Off / 50
Off / 0.1
Off / 600
Off / 600
Off / 35
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Max.
Resolution
500
10
1200
10
10,000
50
25
0.1
4800
50
4800
50
39
0.5
Unit
ml/min
ml/h
ml
ml/h
ml/h
ml/h
°C