4.6

CRRT treatments with citrate anticoagulation

4.6.1

Starting conditions

Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Warning
Any treatment in connection with citrate anticoagulation may only be
performed in intensive care units or under similarly close monitoring.
General description of the Ci-Ca CVVHD and Ci-Ca postCVVHDF
treatments with information on the differences between the individual
treatments.
Make the device ready for operation (see chapter 4.4 on page 4-12).
Note
The status bar shows "CVVHD Ci-Ca" or "CVVHDF Ci-Ca Post" as
treatment mode. If the Ci-Ca module is selected, it is shown in white
characters.
If the Ci-Ca module is deselected, Ci-Ca is shown in grey.
Warning
Please be absolutely sure that the citrate and / or calcium
concentrations as well as the citrate and / or calcium volume of the
solutions used comply with the settings in the Setup menu of the
multiFiltrate.
The concentration of the calcium solution used and of the
calcium-containing substitution fluid must be set correctly in the
System parameters / Select Ci-Ca data menu.
Note
Each treatment mode has its individual starting conditions.
The operator must ensure that the starting conditions are met before
using the rotary selector to select Conditions fulfilled and pressing
[OK].
Chapter 4: Operation
CRRT treatments with citrate anticoagulation
4-33