Fresenius Medical Care multiFiltrate Instructions For Use Manual page 126

Table of Contents

Advertisement

Chapter 4: Operation
Slow continuous ultrafiltration (SCUF)
4.9.5.2
Rinsing the tubing system / entering treatment parameters
4-84
Note
The heparin dose is to be administered as prescribed by the physician!
If an initial heparin dose is to be administered, the bolus function can be
used.
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
The ultrasonic sensor detects fluid in the
venous bubble catcher. Set the venous level
manually using the Venous bubble catcher
level menu field.
Rinsing starts automatically and the screen
for entering the treatment parameters
appears.
Indication of the decreasing rinse
volume and the remaining rinse time.
 Use the rotary selector to select the
required treatment parameters (green
background) and press [OK].
 Use the rotary selector to enter the
required parameters and press [OK].
Set all treatment parameters as described
above.
 Use the rotary selector to select All
treatment parameters entered? [OK] to
confirm! and press [OK].

Advertisement

Table of Contents
loading

Table of Contents