2.12.5 Warnings relating to consumables and accessories ..............2-15 2.13 Addresses ..........................2-16 Design Views of the device ........................3-1 3.1.1 Front view ............................ 3-1 3.1.2 Rear view............................. 3-2 Controls and indicators ......................3-4 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 5
Priming the tubing system ..................4-55 4.7.6.2 Rinsing the tubing system / entering treatment parameters........4-56 4.7.6.3 UF rinse........................4-57 4.7.6.4 Recirculation / waiting for patient ................4-57 4.7.6.5 Connecting the patient ....................4-58 4.7.7 Treatment ..........................4-59 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 6
Recirculation / waiting for patient ................4-85 4.9.5.5 Connecting the patient ....................4-86 4.9.6 Treatment ..........................4-87 4.9.6.1 Treatment main screen ....................4-87 4.9.6.2 Treatment menu ......................4-88 4.9.6.3 Treatment parameters ....................4-88 4.9.7 End of treatment ........................4-89 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 7
Default treatment settings......................4-120 4.12.2 Setting the date / time......................4-120 4.13 Interrupting the treatment..................... 4-121 4.13.1 Disconnecting the patient / interrupting the treatment ............. 4-122 4.13.2 Connecting the patient / continuing the treatment ..............4-123 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 8
Haemoperfusion (HP) ........................ 7-15 Anticoagulation........................7-16 7.3.1 Systemic anticoagulation ......................7-16 7.3.2 Regional anticoagulation with citrate ..................7-17 7.3.2.1 General information on citrate anticoagulation ............7-17 7.3.2.2 Adequate anticoagulation in the extracorporeal blood circuit ........7-22 viii Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 9
Information for the responsible organisation ................10-4 10.4.2 Information for recycling and waste disposal facilities............... 10-4 11 Technical Safety Checks / maintenance procedures 11.1 Important information on the Technical Safety Checks / maintenance procedures..11-1 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 10
Requirements ..........................14-3 14.2.2 Treatment with patient / case ID ....................14-3 14.2.2.1 Previous treatment, continuing ................... 14-3 14.2.2.2 Entering the patient / case ID in UF rinse..............14-4 15 Appendix 15.1 Network............................. 15-1 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 11
15.2 Instructions on the use of "free software"................15-2 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 12
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 16
Chapter 1: Index Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 17
2013 – The page identification, for example, 1-3 refers to chapter 1, page 3. Organisation of the To facilitate the use of documents from Fresenius Medical Care, the chapters organisation of the chapters has been standardised in all manuals.
Advises the operator that failure to observe this information can result in the following: – Damage to the device. – Required functions will not be executed at all or will be executed incorrectly. Significance of tips Information providing useful tips for easy handling. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
(e.g., delivery rates, fill volumes). – Operation in suitable rooms in professional health care facilities. Normative and local regulations must be observed. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
– Patients with whom the haemofilter rapidly and repeatedly becomes clogged when different anticoagulation methods are used. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
CRRT solution can have a more pronounced effect. Regional anticoagulation – Metabolic acid-base disorders (acidosis, alkalosis) with citrate – Systemic hypocalcemia or hypercalcemia – Hypomagnesaemia – Hypernatraemia – Side effects caused by a disordered citrate metabolism Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Note Any treatment in connection with citrate anticoagulation may only be performed in intensive care units or under similarly close medical monitoring. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
2.6.11 Target group The device may only be installed, operated, and used by individuals with the appropriate training, knowledge, and experience and for whom proof of instruction can be shown. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Instructions for Use. The device may only be operated by individuals certificated to have been instructed on the proper operation and handling of the device. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
– The actual values displayed must be compared with the prescribed target values. The device may only be operated under the operating conditions specified by the manufacturer. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
2.12.2 Warnings related to therapy Warning Risk of cross-contamination if tubing systems without hydrophobic filters are used Only use tubing systems with hydrophobic filters in the pressure lines. 2-10 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 27
If you need to move any part of the tubing system out of position, make sure the correct layout of the entire tubing system is restored before continuing treatment, paying special attention to the correct placement of the positioners. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 2-11...
Page 28
Warning Risk of injury due to damaged tubing systems Risk of infection Install the tubing systems in such a way that they cannot be damaged by objects with sharp edges or pets. 2-12 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Danger in case of excessive load (respect the maximum load) If the load on the IV pole is too high, the device may tilt. Ensure that the maximum allowed load on the IV pole of 6 kg is not exceeded. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 2-13...
Please address any queries to the local service support organisation. 2.12.5 Warnings relating to consumables and accessories Warning Consumables must be discarded after the treatment in compliance with the regulations for the disposal of potentially contaminated materials. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 2-15...
Chapter 2: Important information 2.13 Addresses Manufacturer Fresenius Medical Care AG & Co. KGaA 61346 Bad Homburg Germany Phone: +49 6172 609-0 www.fmc-ag.com International Fresenius Medical Care service support Deutschland GmbH Service Support International Hafenstrasse 9 97424 Schweinfurt Germany Phone: +49 9721 678-333 (hotline)
Chapter 3: Design Design Views of the device 3.1.1 Front view Legend IV pole Scale 1 Scale 2 Monitor Extracorporeal blood circuit module Scales 3 and 4 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
3.2.1 Monitor Legend Screen Tube holders Rotary selector and [OK] key [ESC] key [Mute] key [STOP] key [START/RESET] key [On/Off] key Status indicator Green LED (operation) Yellow LED (warning / preparation) Red LED (alarm) Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
The key moves the slide carriage up as long as the key is pressed. The key moves the slide carriage automatically down. Pressing this key again will stop the slide carriage. Syringe holder Slide carriage Grip handle Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
(see chapter 8.1.1 on page 8-2) Filtrate pressure port (yellow) Air detector Blood pump (red dot) Pressure port before filter (white) Arterial pressure port (red) Venous pressure port (blue) Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Chapter 3: Design 3.2.4 multiFiltrate with Ci-Ca module (option) Legend Second IV pole with connector in the IV pole bearing Ci-Ca module Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Citrate insertion switch Calcium drop counter Calcium pump Calcium insertion switch 3.2.4.2 Ci-Ca module rear view Legend Holder for IV pole Securing lever with pressure roll Voltage supply and signal connection to the multiFiltrate Earthing contact Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Notes (white), warnings (yellow), alarms (red), and pressure graphs Parameter fields Confirmation field Screen identification number Menu bar Shows the available menus. The currently selected menu is displayed with black characters and green background. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 42
Chapter 3: Design 3-10 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
– The different pressure conditions on the dialysate and the blood side. – The viscosity of the blood. Warning Aseptic techniques Use aseptic techniques for all blood-side connections and all connections in the area where sterile solutions are to be used. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 44
Ensure that the maximum allowed load on the IV pole of 6 kg is not exceeded. Warning Tubing systems and multiFiltrate kits must be replaced no later than the service life date indicated by the manufacturer. The usable life is specified on the packaging of the tubing systems.
Page 45
If you need to move any part of the tubing system out of position, make sure the correct layout of the entire tubing system is restored before continuing treatment, paying special attention to the correct placement of the positioners. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 46
Warning Citrate and calcium solutions The citrate and calcium solutions must be at room temperature before they can be used. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Pressing the [ESC] key exits the current menu field and activates the menu bar. The selected menu has a green background. Menus that can be selected are shown with white characters on a blue background. Menus that cannot be selected are shown in grey. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
This screen is displayed during the treatment. If another menu function is selected and entered by pressing the [ESC] key whilst a treatment is in progress, pressing the [ESC] key twice will return you to the main screen. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
While changing the value, the actual value will still be displayed in black. The new target value is shown in white below the actual value. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 50
After the new value has been stored by the system, it will be displayed in black. To enter all other parameters, proceed as described above. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
The parameter UF goal can only be entered in the Treatment parameters menu. If ultrafiltration has not been programmed, no UF goal (grey background) can be entered. If ultrafiltration has been programmed, a UF goal (blue background) can be entered. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Resetting the alarm limits in alarm-free condition Press the [START/RESET] key. Press the [START/RESET] key again. The alarm limits will be set around the current actual value. Exception: Venous pressure (see chapter 5.5 on page 5-4) 4-10 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
The bolus anticoagulation function may be selected both during preparation and during the treatment via the Bolus menu field. When changing from preparation to treatment, the bolus quantity will be set to 0. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-11...
(dimmed). If the device is switched on via the "On / Off" key, the green LED goes out and the yellow LED is illuminated. At the end of the functional test, an audible alarm sounds and the red LED is illuminated and then changes back to yellow. 4-12 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 55
After turning the device off, wait for approximately 10 seconds before turning it on again. Connect the device to the power supply. Connect the equipotential bonding connector to the rear of the multiFiltrate device. Press the power switch on the rear of the device.
Starting the functional test Note If the multiFiltrate device is provided with a Ci-Ca module, this module must also pass the functional test after power-on, irrespective of the treatment mode to be selected. If the module fails to pass the functional test even though faults have been eliminated and the functional test has been repeated, treatment with the multiFiltrate device is not permitted.
Note The following screen will not appear if: – The Ci-Ca module is not attached. – Communication between the Ci-Ca module and the multiFiltrate device cannot be established. Contact service support. The decision whether the treatment is to be performed with or without citrate anticoagulation can already be made during the functional test.
Use the rotary selector to select the required treatment mode and press [OK]. The paediatric treatment modes have to be activated by the technician in the service program. Paed. CVVH Paed. CVVHD 4-16 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Insert the cassette system according to the instructions. Turn the rotary selector clockwise until the next screen displays. Continue with "Inserting the heparin syringe" (see chapter 4.5.4 on page 4-21) Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-17...
Turn the rotary selector clockwise until the next screen displays. 4.5.2.3 Inserting the venous blood line system Insert the venous blood line system according to the instructions. Turn the rotary selector clockwise until the next screen displays. 4-18 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
The settings for the tubing system may only be changed by technical service support. With the CVVHD and CVVH (not Pre-Post) treatment modes, both scales can be loaded with solutions if 2 x HF female adapters are used on 4 x HF male adapters. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-19...
Inserting the substituate or sub postdilution tubing system (except for CVVHD) Insert the substituate or sub postdilution tubing system according to the instructions. Turn the rotary selector clockwise until the next screen displays. 4-20 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Blood clots (coagula) can cause the air bubble detector to fail. Note Ensure that the filtrate bag hangs freely and does not touch any other objects. Do not insert the filtrate tube too tightly between the blood leak detector and the filtrate bag. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-21...
Use the rotary selector to select Start priming? [OK] to confirm! and press [OK]. Heparin can be added to the NaCl solution used for priming and rinsing via the Bolus menu field. 4-22 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Use the rotary selector to enter the required parameters and press [OK]. Set all treatment parameters as described above. Use the rotary selector to select All treatment parameters entered? [OK] to confirm! and press [OK]. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-23...
Stop recirculation by pressing the [STOP] key. Press the [STOP] key for approx. 3 seconds. Connect the venous patient line to the arterial patient line using the recirculator. Continue recirculation by pressing the [START/RESET] key. 4-24 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Use the rotary selector to select Start connection? [OK] to confirm! and press [OK]. The blood pump will deliver at the programmed rate. Settable rate: Cassette / adults 10 ml/min to 100 ml/min (default 100 ml/min) Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-25...
The current flow rates (dialysate, substituate, sub predilution, sub postdilution, blood flow, ultrafiltration) Heparin The status bar shows: The treatment mode The progression of the treatment time The balance Continuous anticoagulation on / off 4-26 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Use the rotary selector to select Treatment parameters from the menu bar and press [OK]. The treatment parameters shown depend on the selected treatment mode. After a timeout, the display will automatically return to the treatment main screen. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-27...
Change of procedure to CVVH Use the rotary selector to select Dialysate and set it to Off. 4.5.8.2 Change of procedure to CVVHD Note "Sub. bolus 100 ml" is not possible in CVVHD. 4-28 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Substituate or dialysate must remain deactivated or the end of the treatment has to be initiated. Use the rotary selector to select Substituate or Dialysate and set the required flow. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-29...
[OK] to confirm! and press [OK]. When selecting Disconnect without blood return, the program will go directly to "Disconnecting the patient" without reinfusion after a safety prompt (see chapter 4.5.11 on page 4-32). 4-30 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 73
Use the rotary selector to select Terminate reinfusion and press [OK]. The treatment will be terminated automatically if the reinfusion volume is 0 ml and the optical detector senses non-opaque fluid. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-31...
[OK] to confirm! Confirm with [OK]. 4.5.12 Treatment history Indication of the treatment parameters for the respective treatment mode for the entire treatment. Press the [I/O] key to turn the device off. 4-32 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
System parameters / Select Ci-Ca data menu. Note Each treatment mode has its individual starting conditions. The operator must ensure that the starting conditions are met before using the rotary selector to select Conditions fulfilled and pressing [OK]. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-33...
The following requirements must be fulfilled before inserting the Ci-Ca pump segments to be able to perform automatic priming: – The clamps (white and green) are moved to the connectors. – The clamps (white and green) are closed. 4-34 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
The settings for the tubing system may only be changed by technical service support. 4.6.3.1 Inserting the dialysate tubing system Warning Ci-Ca CVVHD and Ci-Ca postCVVHDF treatments may only be performed if calcium-free dialysate is used. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-35...
For a Ci-Ca postCVVHDF treatment, the substituate line must always be connected to the postdilution port. Insert the substituate tubing system according to the instructions. Turn the rotary selector clockwise until the next screen displays. 4-36 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Blood clots (coagula) can cause the air bubble detector to fail. Note Ensure that the filtrate bag hangs freely and does not touch any other objects. Do not insert the filtrate tube too tightly between the blood leak detector and the filtrate bag. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-37...
15 mm. If this fails, manually readjust the level. Use the rotary selector to select Ci-Ca clamps opened? Ci-Ca drip chambers filled? [OK] to confirm! and press [OK]. 4-38 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Once the Ci-Ca lines are primed while being free from air, use the rotary selector to select Ci-Ca lines primed and free from air? [OK] to confirm! and press [OK]. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-39...
Please ensure that the ratios of blood flow to dialysate flow / substituate flow are properly set. Note Please set the treatment parameters (citrate dose, calcium dose, blood flow, dialysate flow) as described in chapter 7. 4-40 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 83
Use the rotary selector to enter the required parameters and press [OK]. Set all treatment parameters as described above. Use the rotary selector to select All treatment parameters entered? [OK] to confirm! and press [OK]. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-41...
Stop recirculation by pressing the [STOP] key. Press the [STOP] key for approx. 3 seconds. Connect the venous patient line to the arterial patient line using the recirculator. Continue recirculation by pressing the [START/RESET] key. 4-42 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
10 ml/min to 100 ml/min (default 100 ml/min) The citrate pump will deliver at the corresponding rate. It will stop if the optical detector senses opaque fluid or, for safety reasons, after 10 minutes at the latest. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-43...
Warning If the measured post-filter calcium value or systemic calcium value varies significantly, please consult a physician. Observe the instructions on taking a sample (see chapter 7.3.2 on page 7-17). 4-44 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Citrate / blood dose and citrate flow Calcium / filtrate dose and calcium flow Heparin The status bar shows: The treatment mode The progression of the treatment time The balance Continuous anticoagulation on / off Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-45...
Use the rotary selector to select Treatment parameters from the menu bar and press [OK]. The treatment parameters shown depend on the selected treatment mode. After a timeout, the display will automatically return to the treatment main screen. 4-46 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
[OK] to confirm! and press [OK]. When selecting Disconnect without blood return, the program will go directly to "Disconnecting the patient" without reinfusion after a safety prompt (see chapter 4.6.9 on page 4-48). Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-47...
It is forbidden to remove the Ci-Ca lines manually before the patient is disconnected. Warning Consumables must be discarded after the treatment in compliance with the regulations for the disposal of potentially contaminated materials. 4-48 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 91
Remove and dispose of the tubing system. To remove the remaining pump segments, proceed as described above. Use the rotary selector to select View treatment history? [OK] to confirm! Confirm with [OK]. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-49...
Chapter 4: Operation CRRT treatments with citrate anticoagulation 4.6.10 Treatment history Indication of the treatment parameters for the entire treatment. Press the [I/O] key to turn the device off. 4-50 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Make the device ready for operation (see chapter 4.4 on page 4-12). Note The multiFiltrate paed CRRT / SCUF set tubing system has to be used for paediatric CVVH and CVVHD treatment modes. The status bar shows "Paed. CVVHD" or "Paed. CVVH" as the treatment mode.
Turn the rotary selector clockwise until the next screen displays. 4.7.3 Inserting further tubing systems Depending on the selected treatment mode, the tubing system for dialysate or substituate has to be inserted. 4-52 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
4.7.3.2 Inserting the substituate tubing system (except for CVVHD) Insert the substituate tubing system according to the instructions. Turn the rotary selector clockwise until the next screen displays. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-53...
Blood clots (coagula) can cause the air bubble detector to fail. Note Ensure that the filtrate bag hangs freely and does not touch any other objects. Do not insert the filtrate tube too tightly between the blood leak detector and the filtrate bag. 4-54 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Use the rotary selector to select Start priming? [OK] to confirm! and press [OK]. Heparin can be added to the NaCl solution used for priming and rinsing via the Bolus menu field. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-55...
Use the rotary selector to enter the required parameters and press [OK]. Set all treatment parameters as described above. Use the rotary selector to select All treatment parameters entered? [OK] to confirm! and press [OK]. 4-56 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Stop recirculation by pressing the [STOP] key. Press the [STOP] key for approx. 3 seconds. Connect the venous patient line to the arterial patient line using the recirculator. Continue recirculation by pressing the [START/RESET] key. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-57...
Use the rotary selector to select Start connection? [OK] to confirm! and press [OK]. The blood pump will deliver at the programmed rate. Settable rate: Paediatric 10 ml/min to 50 ml/min (default 50 ml/min) 4-58 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
The current flow rates (dialysate, substituate, sub predilution, sub postdilution, blood flow, ultrafiltration) Heparin The status bar shows: The treatment mode The progression of the treatment time The balance Continuous anticoagulation on / off Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-59...
Use the rotary selector to select Treatment parameters from the menu bar and press [OK]. The treatment parameters shown depend on the selected treatment mode. After a timeout, the display will automatically return to the treatment main screen. 4-60 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
[OK] to confirm! and press [OK]. When selecting Disconnect without blood return, the program will go directly to "Disconnecting the patient" without reinfusion after a safety prompt (see chapter 4.7.9 on page 4-63). Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-61...
Page 104
Use the rotary selector to select Terminate reinfusion and press [OK]. The treatment will be terminated automatically if the reinfusion volume is 0 ml and the optical detector senses non-opaque fluid. 4-62 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Confirm with [OK]. 4.7.10 Treatment history Indication of the treatment parameters for the entire treatment. The menu fields shown depend on the selected treatment mode. Press the [I/O] key to turn the device off. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-63...
Insert the cassette system according to the instructions. Turn the rotary selector clockwise until the next screen displays. Continue with "Inserting the heparin syringe" (see chapter 4.8.4 on page 4-67) 4-64 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Turn the rotary selector clockwise until the next screen displays. 4.8.2.3 Inserting the venous blood line system Insert the venous blood line system according to the instructions. Turn the rotary selector clockwise until the next screen displays. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-65...
Turn the rotary selector clockwise until the next screen displays. 4.8.3 Inserting the plasma line system Insert the plasma line system according to the instructions. Turn the rotary selector clockwise until the next screen displays. 4-66 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Blood clots (coagula) can cause the air bubble detector to fail. Note Ensure that the filtrate bag hangs freely and does not touch any other objects. Do not insert the filtrate tube too tightly between the blood leak detector and the filtrate bag. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-67...
Use the rotary selector to select Start priming? [OK] to confirm! and press [OK]. Heparin can be added to the NaCl solution used for priming and rinsing via the Bolus menu field. 4-68 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Use the rotary selector to enter the required parameters and press [OK]. Set all treatment parameters as described above. Use the rotary selector to select All treatment parameters entered? [OK] to confirm! and press [OK]. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-69...
Stop recirculation by pressing the [STOP] key. Press the [STOP] key for approx. 3 seconds. Connect the venous patient line to the arterial patient line using the recirculator. Continue recirculation by pressing the [START/RESET] key. 4-70 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
[OK] to confirm! and press [OK]. The blood pump will deliver at the programmed rate. Settable rate: 10 ml/min to 100 ml/min (default 100 ml/min) Time-controlled filling of plasma is in progress. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-71...
Disconnect the venous patient line from the NaCl solution and connect it to the venous vascular access. Use the rotary selector to select Start treatment? [OK] to confirm! and press [OK]. 4-72 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Detailed description (see chapter 4.11 on page 4-104). Deactivate the blood leak detector (see chapter 5.13 on page 5-16). After a timeout, the display will automatically return to the treatment main screen. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-73...
The scales alert the operator that the plasma bag will need to be replaced in 5 minutes. Press the [ESC] key. Use the rotary selector to select Treatment from the menu bar and press [OK]. 4-74 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 117
[OK]. It is imperative to enter the new net volume of the replacement solution. Use the rotary selector to select Terminate bag change? [OK] to confirm! and press [OK]. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-75...
[OK]. Use the rotary selector to select Start plasma return? [OK] to confirm! and press [OK]. The residual plasma of approx. 300 ml will be administered until the Balancing message is displayed. 4-76 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
[OK] to confirm! and press [OK]. When selecting Disconnect without blood return, the program will go directly to "Disconnecting the patient" without reinfusion after a safety prompt (see chapter 4.8.9 on page 4-79). Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-77...
Page 120
Use the rotary selector to select Terminate reinfusion and press [OK]. The treatment will be terminated automatically if the reinfusion volume is 0 ml and the optical detector senses non-opaque fluid. 4-78 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Use the rotary selector to select View treatment history? [OK] to confirm! Confirm with [OK]. 4.8.10 Treatment history Indication of the treatment parameters for the entire treatment. Press the [I/O] key to turn the device off. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-79...
Slow continuous ultrafiltration (SCUF) Prepare the device ready for operation (see chapter 4.4 on page 4-12). Note The multiFiltrate paed CRRT / SCUF set tubing system (part no. 501 775 1) has to be used for the SCUF treatment mode. 4.9.1 Starting conditions Note Each treatment mode has its individual starting conditions.
Turn the rotary selector clockwise until the next screen displays. 4.9.2.3 Inserting the filtrate line system Insert the filtrate line system according to the instructions. Turn the rotary selector clockwise until the next screen displays. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-81...
Page 124
Blood clots (coagula) can cause the air bubble detector to fail. Note Ensure that the filtrate bag hangs freely and does not touch any other objects. Do not insert the filtrate tube too tightly between the blood leak detector and the filtrate bag. 4-82 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Use the rotary selector to select Start priming? [OK] to confirm! and press [OK]. Heparin can be added to the NaCl solution used for priming and rinsing via the Bolus menu field. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-83...
Page 126
Use the rotary selector to enter the required parameters and press [OK]. Set all treatment parameters as described above. Use the rotary selector to select All treatment parameters entered? [OK] to confirm! and press [OK]. 4-84 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Stop recirculation by pressing the [STOP] key. Press the [STOP] key for approx. 3 seconds. Connect the venous patient line to the arterial patient line using the recirculator. Continue recirculation by pressing the [START/RESET] key. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-85...
Page 128
Use the rotary selector to select Start connection? [OK] to confirm! and press [OK]. The blood pump will deliver at the programmed rate. Settable rate: Cassette / adults 10 ml/min to 100 ml/min (default 100 ml/min) 4-86 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
The current pressures (arterial, venous, TMP) The current flow rates (blood flow, ultrafiltration) Heparin The status bar shows: The treatment mode The progression of the treatment time The balance Continuous anticoagulation on / off Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-87...
Press the [ESC] key. Use the rotary selector to select Treatment parameters from the menu bar and press [OK]. After a timeout, the display will automatically return to the treatment main screen. 4-88 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
[OK] to confirm! and press [OK]. When selecting Disconnect without blood return, the program will go directly to "Disconnecting the patient" without reinfusion after a safety prompt (see chapter 4.9.8 on page 4-91). Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-89...
Page 132
Use the rotary selector to select Terminate reinfusion and press [OK]. The treatment will be terminated automatically if the reinfusion volume is 0 ml and the optical detector senses non-opaque fluid. 4-90 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 133
Use the rotary selector to select View treatment history? [OK] to confirm! Confirm with [OK]. 4.9.9 Treatment history Indication of the treatment parameters for the entire treatment. Press the [I/O] key to turn the device off. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-91...
Page 134
Insert the cassette system according to the instructions. Turn the rotary selector clockwise until the next screen displays. Continue with "Inserting the heparin syringe" (see chapter 4.10.3 on page 4-94) 4-92 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 135
Turn the rotary selector clockwise until the next screen displays. 4.10.2.3 Inserting the venous blood line system Insert the venous blood line system according to the instructions. Turn the rotary selector clockwise until the next screen displays. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-93...
Page 136
4.10.4 Complete tubing arrangement Warning Patient hazard: air embolism The air detector must be clean and dry. Do not use any ultrasound-conducting objects or media. Blood clots (coagula) can cause the air bubble detector to fail. 4-94 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 137
Use the rotary selector to select Start priming? [OK] to confirm! and press [OK]. Heparin can be added to the NaCl solution used for priming and rinsing via the Bolus menu field. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-95...
Page 138
Use the rotary selector to enter the required parameters and press [OK]. Set all treatment parameters as described above. Use the rotary selector to select All treatment parameters entered? [OK] to confirm! and press [OK]. 4-96 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 139
NaCl solution. If using an NaCl solution with one connector: Connect the venous patient line to the arterial patient line using the recirculator. Continue recirculation by pressing the [START/RESET] key. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-97...
Page 140
Use the rotary selector to select Start connection? [OK] to confirm! and press [OK]. The blood pump will deliver at the programmed rate. Settable rate: Cassette / adults 10 ml/min to 100 ml/min (default 100 ml/min) 4-98 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 141
The current pressures (arterial, venous, pF) The current flow rates (blood flow) Heparin The status bar shows: The treatment mode The progression of the treatment time Continuous anticoagulation on / off Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-99...
Page 142
Press the [ESC] key. Use the rotary selector to select Treatment parameters from the menu bar and press [OK]. After a timeout, the display will automatically return to the treatment main screen. 4-100 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
[OK] to confirm! and press [OK]. When selecting Disconnect without blood return, the program will go directly to "Disconnecting the patient" without reinfusion after a safety prompt (see chapter 4.10.8 on page 4-103). Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-101...
Page 144
Use the rotary selector to select Terminate reinfusion and press [OK]. The treatment will be terminated automatically if the reinfusion volume is 0 ml and the optical detector senses non-opaque fluid. 4-102 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Use the rotary selector to select View treatment history? [OK] to confirm! Confirm with [OK]. 4.10.9 Treatment history Indication of the treatment parameters for the entire treatment. Press the [I/O] key to turn the device off. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-103...
/ performed if a calcium-containing HF solution / dialysate is used. Warning After deselecting citrate anticoagulation, the Ci-Ca lines must not be removed from the pumps before the patient has been completely disconnected. 4-104 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
When starting the treatment, define the post-filter calcium value. If the ionised calcium has not decreased at this point, it is absolutely necessary that the tubing system and the solutions used are checked. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-105...
Page 148
[OK]. Use the rotary selector to select All conditions fulfilled? [OK] to confirm! and press [OK]. After selecting citrate anticoagulation, the multiFiltrate immediately switches to the menu for dialysate bag change. 4-106 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
[OK]. 4.11.4 Ci-Ca bag change Note The Ci-Ca bag change menu can be accessed: – Manually by selecting the Ci-Ca bag change menu. – Automatically after a bag change notification. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-107...
Page 150
Select Cancel bag change to return to the Treatment menu. Use the rotary selector to select Citrate bag change completed or Ca bag change completed from the menu field and press [OK]. 4-108 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Note on C The syringe wings must be placed in the syringe wing slot. Note on D – The thumbplate of the syringe plunger must be positioned between the clamps of the grip handle. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-109...
The balance is the amount of fluid that is removed from and added to the patient during the treatment. During the CRRT treatments and SCUF, all volumes (substituate, citrate, calcium, heparin) which are added continuously are completely removed by the filtrate pump. 4-110 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
"zero". The treatment time and the filter life will not be reset. – Balance data development – Balance previous treatment Display: – Treatment time – Filter Life Further values displayed: – Dependent on the treatment mode Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-111...
Use the rotary selector to select Start time and End time and press [OK]. Enter times and press [OK] to confirm. The balance data of the period of time entered is calculated automatically and displayed. 4-112 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Treatment menu 4.11.7.5 Total balance after the treatment Indication of the treatment parameters for the entire treatment. 4.11.7.6 Balance since reset The balance data since the last balance data reset will be displayed. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-113...
The absolute lower alarm limit for venous pressure during treatment is +10 mmHg. The lower pressure limit is adjusted accordingly. Following this, the pressure limits should be checked by the operator and adjusted individually as required. 4-114 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Use the rotary selector to enter the required window width and press [OK]. 4.11.8.3 Exiting the Alarm limits menu Use the rotary selector to select Exit Alarm limits menu? [OK] to confirm! and press [OK]. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-115...
Use the rotary selector to select Page and press [OK] to confirm. Ten events per page will be displayed when you turn the rotary selector. 4-116 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
The bolus will be included in the calculation of the balance. Sub. bolus cannot be administered in "Balancing off" and the following treatment modes: – Ci-Ca CVVHD – CVVHD – HP – SCUF – MPS – Paed. CVVHD and Paed. CVVH Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-117...
Page 160
Treatment menu The upper venous alarm limit is set to 300 mm/Hg. The lower TMP alarm limit is set to –60 mm/Hg. The bolus can be stopped at any time by selecting Stop bolus. 4-118 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Tubing arrangement When Detailed is selected, the installation of the line system will be shown step by step. When Complete is selected, the fully inserted line system will be displayed. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-119...
The parameters that were changed are not activated until the device is brought back into operation with the [I/O] key. 4.12.2 Setting the date / time Set the current date and time. 4-120 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
– The stopped anticoagulation process must be restarted. Warning Always observe aseptic procedures when connecting and disconnecting the patient and perform the process with due care to avoid any loss of blood. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-121...
When the optical detector senses non-opaque fluid, disconnect the venous patient line from the patient and connect it to the NaCl solution. Press the [START/RESET] key. Balancing stops automatically. Ci-Ca pumps stop automatically. 4-122 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
NaCl solution and connect it to the venous vascular access. Continue the treatment by pressing the [START/RESET] key. Balancing starts automatically. Ci-Ca pumps start automatically. If necessary, set blood flow and heparin administration. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-123...
Page 166
Chapter 4: Operation Interrupting the treatment 4-124 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Wrong dosage of anticoagulation medium Blood leak alarms Loss of blood into filtrate / plasma Isolated citrate dosage with Citrate accumulation / balancing switched off disruption of acid-base status Low temperature warning Hypothermia Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
The operating status indicator is yellow and the signal emits an audible warning. Internally, the alarms and warnings are given different priorities (described in the following section). The system also offers the option of using isolated audible signals to inform the operator. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Since critical alarm situations can cause the machine to be placed in Safe Mode (halting treatment), alarms with high priority are only triggered in exceptional cases, where a potential hazardous condition persists even following alarm response. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Full sensitivity at max. blood flow is achieved with patient weights upwards of 45 kg. If return pressure is negative (e.g., when deploying central venous catheters) then the risk of an air infusion is present, even downstream of the air detector. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 171
Audible alarm Silencing the audible alarm will always suppress an alarm for a duration suppression / of 2 minutes. Any new alarm that occurs will activate the silenced mute audible alarm. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
The Mute key allows the operator to suppress (deactivate) the audible tone of a signaled alarm for a certain period of time. If a new alarm occurs during this time, the audible alarm tone of the new alarm is signaled regardless. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
This number will enable the service to provide faster assistance in the event of a problem. Selecting the ? button will display additional support on how to correct the messages. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Balance is off. By selecting the green question mark with the rotary selector and pressing [OK], the operator will obtain support in correcting the cause of the problem. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Balance is off. The blood pump is stopped. By selecting the green question mark with the rotary selector and pressing [OK], the operator will obtain support in correcting the cause of the problem. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Repeat the functional test by pressing the [START/RESET] key. Note If the functional test was not passed even after having been repeated, and if the problem cannot be corrected, call service support. 5-10 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Press the [START/RESET] key. The blood pump will start at a slower speed and gradually increase to the prescribed rate. Then the required balancing pumps will start (depending on the treatment mode). Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 5-11...
20% of the BP rate. Substituate + ultrafiltration = UF rate 2000 ml/h + 200 ml/h = 2200 ml/h = 36.6 ml/min At a BP rate of 150 ml/min, this equals a UFR / BPR of 24.4%. 5-12 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 179
This warning can be confirmed with the [START/RESET] key provided the treatment is performed in predilution or the imbalance in postdilution is required. It is advisable to correct this imbalance by changing the parameters. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 5-13...
Setup menu of the multiFiltrate. The calcium flow is limited by the control range of the calcium pump. Control range of the calcium pump: 10–100 ml/h...
The citrate flow rate is calculated by the system depending on the set citrate dose, the set blood flow, and the concentration of the citrate solution used (set in the multiFiltrate Setup menu), and is limited by the control range of the citrate pump.
It is important to pay attention to additional darkening due to a blood leak in the filtrate bag, particularly when using a permanent override for plasma separation. 5-16 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 183
[Yes], the blood leak detector no longer monitors for haemolysis or blood loss (Blood leak detector deactivated!). Monitoring can be reactivated at any time in the Treatment menu. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 5-17...
When power is restored, the system will start automatically. End of emergency operation is displayed. Balance is off. The blood pump is stopped. The extracorporeal blood volume must be returned using a hand crank (see chapter 5.15 on page 5-19). 5-18 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
When the blood has been returned manually, close the clamps on the tubing system and remove the venous patient line. Dismantle the tubing system manually. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 5-19...
Page 186
Chapter 5: Alarm processing 5-20 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Active substance base: 45% ethanol Type of disinfection: quick-acting disinfectant for wipe disinfection combined with Freka-Wipes disposable cloths ClearSurf Wipes Ready to use wipes Active substance base: cationic surfactants Type of disinfection: wipe disinfection Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 188
Chapter 6: Cleaning / disinfection Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
During a Ci-Ca treatment, a significant amount of calcium is removed from the patient's blood. For this reason, a calcium substituate is infused into the venous return line by the integrated calcium pump. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
20%. If this limit is exceeded, this will be displayed on the device. If the haemoconcentration is too high, it can be reduced by increasing the blood flow or reducing the substitution in postdilution. It is also possible to select another CRRT treatment as an alternative. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 191
Blood flow ml/min Ultrafiltration Off / 10 1800 ml/h UF goal Off / 50 10,000 Cont. heparin adm. Off / 0.1 ml/h Substituate 4800 ml/h Temperature Off / 35 °C Fig.: CVVH flow diagram Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 192
1200 ml/h UF goal Off / 50 10,000 Cont. heparin adm. Off / 0.1 ml/h Sub predilution Off / 600 4800 ml/h Sub postdilution Off / 600 4800 ml/h Temperature Off / 35 °C Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 194
Min. Max. Resolution Unit Blood flow ml/min Ultrafiltration Off / 10 1800 ml/h UF goal Off / 50 10,000 Cont. heparin adm. Off / 0.1 ml/h Dialysate 4800 ml/h Temperature Off / 35 °C Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 195
Combination of the basic These three basic treatments can be combined in pairs: treatments CVVHDF Pre CVVH + CVVHD => Pre CVVHDF Post CVVH + CVVHD => Post CVVHDF Pre CVVH + Post CVVH => Pre-Post CVVH Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 196
(UFR / BPR) is not affected by this. Fig.: CVVHDF flow diagram Legend Blood pump Haemofilter Air detector Venous clamp Filtrate pump Blood leak detector Filtrate Scales 3 and 4 Predilution Postdilution Check valve Heater Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 197
It was revealed that the filter operating lives are significantly longer in CVVHD than in Pre-Post CVVH, even if the limitation of the filtration fraction to 20% in Pre-Post CVVH was observed. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
2 l/h for a typical adult with a body weight of 70 kg. The value 1.73 m² is the general value for the body surface of a typical adult. 7-10 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Replacement solution As the plasma removed contains colloid osmotically active proteins such as albumin, replacement solutions containing iso-oncotic colloids are typically used. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 7-11...
Page 200
MPS treatments will not be performed until the following day at the earliest. Treatment parameters Min. Max. Resolution Unit Blood flow ml/min Plasma Off / 10 ml/min Temperature Off / 35 °C Cont. heparin adm. Off / 0.1 ml/h 7-12 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Treatment parameters Available settings Min. Max. Resolution Unit Blood flow ml/min Cont. heparin adm. Off / 0.1 ml/h Fig.: Schematic of a haemoperfusion treatment (HP) Legend Blood pump Adsorber cartridge Air detector Venous clamp Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 7-15...
This pump can also be used to administer a bolus when required. An infusion line for anticoagulants is included in the tubing system. Fig.: Schematic of systemic anticoagulation Legend Blood pump Haemofilter Air detector Venous clamp Heparin pump 7-16 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Off / 10 14,400 Cont. heparin adm. Off / 0.1 ml/h Dialysate 4800 ml/h Substituate 2400 ml/h Citrate / blood (citrate dose) mmol/l Calcium / filtrate (calcium dose) mmol/l Temperature Off / 35 °C Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 7-17...
Page 206
This is possibly explained by the fact that activated thrombin, as opposed to prothrombin (thrombinogen, factor II), no longer has a GLA domain, which explains why the mechanism discussed above cannot be effective with thrombin. 7-18 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 207
When performing a Ci-Ca treatment, this is checked downstream of the filter. Note It is absolutely necessary to ensure that the two measurements of the ionised calcium are not mistaken for each other. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 7-19...
Page 208
In case of recirculation in the area of the catheter, blood which has just been returned will enter into the aspirated blood, along with some citrate. This addition of citrate leads to incorrect measurement values for the systemic ionised calcium. 7-20 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 209
HF solution / dialysate. Therefore, the Na concentration of Ci-Ca dialysates K2 and Ci-Ca dialysates K4 has been adjusted to 133 mmol/l. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 7-21...
All kits for Ci-Ca treatments permit a representative check after 5 minutes even in case of a low blood flow of 80 ml/min. 7-22 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
In Ci-Ca postCVVHDF, the calcium concentration of the substitution fluid infused in postdilution is also considered in the calculation. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 7-23...
Page 212
A clogged membrane has severely restricted functionality and must be replaced (restart treatment following replacement of the multiFiltrate kit). 7-24 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
One positive extra effect of the increased citrate dose is that it inhibits a premature increase in the concentration of ionised calcium following infusion of the calcium-containing substitution fluid at the postdilution stage. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 7-25...
Page 214
The full effect can, however, only be assessed after approximately one day. This is particularly important to bear in mind if several equivalent changes are made within short intervals as these may cause an excessive response. 7-26 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
After completion of the treatment, the accumulated calcium-citrate after citrate complexes are metabolised by the patient. This may result in alkalosis anticoagulation and hypercalcaemia. If clinically indicated, these risks can be reduced by continuing the CRRT treatment without citrate anticoagulation. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 7-27...
Depending on the citrate and calcium solutions used locally, the concentration of the citrate and calcium ions must be stored in mmol/l in the System parameters menu of the multiFiltrate. This is done by technical service personnel. This also applies to the filling volume of the storage containers used.
Page 217
(corresponding to Ci-Ca dialysate K4 appropriate concentration 136 mmol/l citrate) per 5 litre bag (50 to 500 mmol/l calcium 1.0 litre bag ions); In addition, a preferably approx. calcium-containing, 100 mmol/l bicarbonate-buffered substitution fluid Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 7-29...
Page 218
Chapter 7: Functional description 7-30 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Disposables Identified by the symbol: Do not re-use Use by date Identified by the symbol: Use by Long-term operation Identified by the symbol: Indication of the max. operating time and the max. delivery volume Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Chapter 8: Consumables, accessories, additional equipment Consumables 8.1.1 multiFiltrate kits The ready-to-use multiFiltrate kits comprise all tubing systems required for the respective therapy. Used in conjunction with this device, these tubing systems constitute a defibrillator-proof applied part. Product Part number Description ®...
1.5 l bag 8.1.6 Disposable syringes Product Part number Description Fresenius Medical Care heparin syringe, 503 032 1 30 ml Fresenius Injectomat syringe, 50 ml 900 071 1 Fresenius P syringe, 50 ml 900 076 1 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Hydrophobic filter for connection to male Luer-Lock Check valve 850 080 1 Male / female Luer-Lock Clamp 284 024 1 For clamping tubes Freka-Flex transfer system 288 901 1 Infusion system with roller clamps and drip chamber Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
(RCD less than 30 mA) for each device or electric circuit. – Overvoltage / lightning protection in the main and emergency power supply networks. – A connecting bolt must be available for an additional equipotential bonding conductor. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Documentation The initial start-up report and further procedural information are described in the Service Manual. Reports are available on request. The completion of the initial start-up must be entered in the Medical Device Register. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Leakage currents If additional equipment not listed in the Accessories chapter is connected to the device, there is a danger that the permitted leakage currents will be exceeded. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 228
Chapter 9: Installation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Do not use the handles on the upper part of the unit (monitor hood) to lift the system. 10.1 Relocation Moving Since the system is provided with a cart, the device can easily be relocated (by moving it on a level surface) within a building. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 10-1...
Page 230
The device indicates a power failure. The audible alarm can be suppressed for 2 minutes by pressing the [Mute] key. The device must be relocated as fast as possible, as battery operation is only possible for a limited time. 10-2 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
– Use the power cable to connect the device to the power supply. – Actuate the power switch to turn the device on. – Leave the device on for 10 hours. If the device is not used, repeat this procedure every six months. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 10-3...
Batteries and rechargeable batteries must be properly disposed of in accordance with the applicable local regulations. The device includes electronic circuit boards and an LCD screen. More information will be made available to waste disposal facilities on request. 10-4 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
The Technical Safety Checks, the maintenance points, and further procedural information are described in the Service Manual. Reports are available on request. The completion of the Technical Safety Checks must be entered in the Medical Device Register. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 11-1...
Page 234
Chapter 11: Technical Safety Checks / maintenance procedures 11-2 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Weight with options: approx. 100 kg Safe working load: 55 kg Maximum weight: approx. 155 kg IV pole load-bearing Total IV pole load-bearing capacity: max. 6 kg capacity Hook load-bearing capacity: max. 5 kg Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 12-1...
2: Protection against touch and foreign bodies with a diameter of at least 12.5 mm 1: Protection against ingress of liquids: Drip-proof 12.2.2 Voltage label Legend Connected load Operating current Degree of protection against electric shock 12-2 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Samples of approved tubing systems are listed in chapter 8 (see chapter 8.1.1 on page 8-2). Degree of protection Drip-proof, IP 21 against ingress of liquids Leakage currents According to EN 60601-1 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 12-3...
+24 V +1 V / –2 V Power switch Main power switch, all-pole, simultaneous disconnection Lead-acid battery (maintenance-free), 18 V / ≥3 Ah Battery Operating time approx. 15 minutes Blood circuit only, without balancing, without heater 12-4 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
A device malfunction can be caused by high-frequency electromagnetic waves (radiated interference source). Observe the minimum required distances for radiated interference sources. For further information on "Guidance and manufacturer's declaration on EMC", refer to the Service Manual. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 12-5...
Page 240
If a source of radiated interference uses a radio frequency technology not specified in the table, the required minimum distance to medical electrical devices can be obtained from the manufacturer. 12-6 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Guidance and manufacturer's declaration on electromagnetic emissions The multiFiltrate device is designed for operation in an electromagnetic environment as specified below. The customer or operator of the multiFiltrate device should ensure that it is operated in such an environment. Interference emission Compliance Electromagnetic environment –...
Page 242
Note: U is the AC voltage before the test level is applied. Portable and mobile radio devices should not be any closer to the multiFiltrate device (including the lines) than the recommended safety distance, which is calculated according to the relevant equation for the transmission frequency.
To determine the electromagnetic environment with regard to stationary transmitters, consider carrying out a study of the site. If the measured field strength at the site where the multiFiltrate device is used exceeds the aforementioned compliance level, the multiFiltrate should be observed in order to verify that it is working as intended.
– Updates or upgrades of components and devices in the IT network Note that local laws take priority over the above-mentioned normative requirements. Please address any queries to the local service support organisation. 12-10 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Priming the tubing Automatic priming and deaeration of the tubing systems by pressing a systems key. Priming will stop automatically as soon as the venous bubble catcher is filled. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 12-11...
< 1% related to the total volume delivered during 24 hours of treatment, circuit in horizontal position and with a maximum inclination of 5°. A deviation of ±100 ml is possible if the total volume delivered is low (typically up to 5000 ml). 12-12 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 247
50 ml/min, a temperature of 37 °C will be reached. – The donor plasma (FFP) must be warmed to a minimum temperature of 20 °C prior to the treatment. (* = Important performance characteristics according to IEC 60601-1) Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 12-13...
Accuracy: 10 mmHg Alarm window width: 20 mmHg to 200 mmHg, size and position adjustable around actual pressure The pF is calculated using the following formula and then displayed: pF = P – P 12-14 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 249
Delivery accuracy: pressure-dependent, ±10% (above typical pressure range); for delivery rates < 6 ml/h, the deviation can be ±20% Delivery rate: 10 to 600 ml/h (citrate / blood ratio: citrate concentration per litre of delivered blood: 2–6 mmol/l) Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 12-15...
The dose is specified in mmol per litre of filtrate (filtrate flow). Calcium flow The calcium flow is the volume of calcium solution supplied to the patient’s blood per time unit. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 13-1...
Page 254
Convection Convection describes the transport of solutes together with the solvent (drag effect, e.g., haemofiltration). ME device Medical electrical device. Non-ME device Non-medical electrical device. Postdilution Adding substituate downstream of the haemofilter. 13-2 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Degree of protection against electric shock: Defibrillator-protected applied part, type BF Degree of protection against electric shock: Defibrillator-protected applied part, type CF Alternating current Equipotential bonding connection Risk of tilting when inclined by more than 5° 13-4 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 257
Serial number Maximum total weight Follow Instructions for Use Warning – Risk of tilting when pushing the device or leaning against it General warning sign Warning – Excessive weight load (observe the maximum load) Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 13-5...
Catalog number Temperature limitation +30°C +5°C Pump segment diameter Sterile. Sterilised using ethylene oxide STERILE EO Sterile. Sterilised using irradiation STERILE R Sterile. Sterilised using steam STERILE Max. service life and max. delivery quantity 13-6 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Chapter 13: Definitions 13.5 Certificates The local service support organisation can provide the currently applicable version of the EC certificates on request. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 13-7...
Page 260
Chapter 13: Definitions 13-8 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Ci-Ca module has been completed successfully. – Is completed as soon as the optical detector of the multiFiltrate system senses opaque fluid. Citrate anticoagulation –...
Page 262
If citrate bag change is selected An alarm will be emitted if bag change is not completed within 2 minutes of selecting the citrate bag change. This alarm will be repeated every 2 minutes. 14-2 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
The Cat. 5 LAN connection cables listed in the following are designed to connect the multiFiltrate to a network hub or switch. Direct connection of the acute dialysis device to a PC requires a Cat. 5 crossover patch cable.
Case ID and press [OK] in order to change the patient / case ID. To confirm the patient / case ID, use the rotary selector to choose Patient/case ID correct? [OK] to confirm! and press [OK]. 14-4 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 265
Accept ID? [OK] to confirm! and press [OK]. After entering and checking the patient ID and / or the case ID, use the rotary selector to select Patient/case ID correct? [OK] to confirm! and press [OK]. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 14-5...
Page 266
Chapter 14: Options 14-6 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Note For reasons of data protection, the operator is responsible for the safekeeping of the PatientCard. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 15-1...
Page 268
/ or provides the source code of such free software. Fresenius Medical Care meets these requirements by providing the copyright notices, remarks, and license texts contained in section C below.
For that reason, section B contains an appropriate note. In such a case, Fresenius Medical Care will stop any further technical support for the device involved. In addition, such modifications and / or manipulations may result in the voiding of warranty claims against Fresenius Medical Care or other vendors of the haemodialysis system if a claim has arisen or might arise in respect thereto.
Page 270
The precise terms and conditions for copying, distribution and modification follow. GNU GENERAL PUBLIC LICENSE TERMS AND CONDITIONS FOR COPYING, DISTRIBUTION AND MODIFICATION 15-4 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 271
Sections 1 and 2 above provided that you also do one of the following: a) Accompany it with the complete corresponding machine-readable source code, which must be distributed under the terms of Sections 1 and 2 above on a medium customarily used for software interchange; or, Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 15-5...
Page 272
This section is intended to make thoroughly clear what is believed to be a consequence of the rest of this License. 15-6 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
If you want us to send this CD to you, please inform us accordingly, by e-mail, telefax, or mail, under the address given in the Instructions for Use. Please do not forget to specify the system type and the system number. II. GNU Lesser General Public License, version 2.1 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 15-7...
Page 274
15-8 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 275
"work based on the library" and a "work that uses the library". The former contains code derived from the library, whereas the latter must be combined with the library in order to run. GNU LESSER GENERAL PUBLIC LICENSE TERMS AND CONDITIONS FOR COPYING, DISTRIBUTION AND MODIFICATION Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 15-9...
Page 276
Library, the distribution of the whole must be on the terms of this License, whose permissions for other licensees extend to the entire whole, and thus to each and every part regardless of who wrote it. 15-10 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 277
Library among them, as well as a reference directing the user to the copy of this License. Also, you must do one of these things: Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 15-11...
Page 278
Library subject to these terms and conditions. You may not impose any further restrictions on the recipients' exercise of the rights granted herein. You are not responsible for enforcing compliance by third parties with this License. 15-12 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...
Page 279
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. THE ENTIRE RISK AS TO THE QUALITY AND PERFORMANCE OF THE LIBRARY IS WITH YOU. SHOULD THE LIBRARY PROVE DEFECTIVE, YOU ASSUME THE COST OF ALL NECESSARY SERVICING, REPAIR OR CORRECTION. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 15-13...
Software without prior written authorisation from him. V. Catharon License (Libft2lib) Part of this software is based on programmings made by Catharon Productions, Inc. Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 15-15...
Page 282
Chapter 15: Appendix 15-16 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015...