Chapter 4: Operation
4.4

Making the device ready for operation

4.4.1

Turning the device on

4-12
Note
Check the device for stability. The maximum angle of inclination must
not exceed 5°.
Before turning the device on, please ensure that all scales are
unloaded, that no objects are placed on the scales and that no tubing
system is inserted in the Ci-Ca pumps.
The maximum loading capacity of 12 kg per scale must not be
exceeded. The weighing cell can even be permanently damaged by a
short-term overload (e.g., pulling or lifting the device by the scales), in
which case the device can no longer be used.
Note
The operator must check all of the status indicator lights.
Once mains power is switched on, the green LED is illuminated
(dimmed). If the device is switched on via the "On / Off" key, the green
LED goes out and the yellow LED is illuminated. At the end of the
functional test, an audible alarm sounds and the red LED is illuminated
and then changes back to yellow.
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
 Use the rotary selector to select Bolus
and press and hold [OK].
As long as the selector is being pressed,
the bolus is slowly administered at a rate
of 0 ml up to 5 ml maximum (adjustable
via the System parameters menu). The
bolus will stop as soon as the selector is
released.
If the bolus, once activated, is stopped before
reaching the maximum volume, bolus
administration may be continued within
5 seconds until reaching the maximum
volume.