Disconnecting The Patient And Removing The Tubing System; Treatment History - Fresenius Medical Care multiFiltrate Instructions For Use Manual

Table of Contents

Advertisement

4.7.9

Disconnecting the patient and removing the tubing system

4.7.10 Treatment history

Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Warning
Consumables must be discarded after the treatment in compliance with
the regulations for the disposal of potentially contaminated materials.
Paediatric CRRT treatments
 Disconnect the venous patient line from
the patient.
 Remove the pump segment adapter from
the groove.
 Press and hold the [START/RESET] key
until the pump segment has been
completely removed.
Support the removal of the pump
segment by slightly pulling on it.
 Remove the heater bag (depending on the
treatment mode) from the heater.
Open the clamps before and after the
heater bag to facilitate removal of the
bag.
 Remove and dispose of the tubing
system.
To remove the remaining pump segments,
proceed as described above.
 Use the rotary selector to select View
treatment history? [OK] to confirm!
Confirm with [OK].
Indication of the treatment parameters for the
entire treatment. The menu fields shown
depend on the selected treatment mode.
 Press the [I/O] key to turn the device off.
Chapter 4: Operation
4-63

Advertisement

Table of Contents
loading

Table of Contents