End Of Treatment; Terminating The Treatment; Starting Reinfusion - Fresenius Medical Care multiFiltrate Instructions For Use Manual

Table of Contents

Advertisement

4.6.8

End of treatment

4.6.8.1

Terminating the treatment

4.6.8.2

Starting reinfusion

Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
CRRT treatments with citrate anticoagulation
 Press the [ESC] key.
 Use the rotary selector to select End of
treatment from the menu bar and press
[OK] to confirm.
The blood pump is running.
Balance is off.
The calcium pump stops.
The citrate pump runs until the optical
detector senses non-opaque fluid, but for no
longer than 10 min.
 Stop the treatment by pressing the
[STOP] key.
Press the [STOP] key for approx.
3 seconds.
The venous clamp closes.
The blood pump is stopped.
or
 Use the [ESC] key to select a different
menu from the menu bar.
 Continue the treatment by pressing the
[START/RESET] key.
or
 Disconnect the arterial patient line from
the patient and connect it to the NaCl
solution.
 Use the rotary selector to select Start
disconnection? [OK] to confirm! and
press [OK].
When selecting Disconnect without blood
return, the program will go directly to
"Disconnecting the patient" without reinfusion
after a safety prompt (see chapter 4.6.9 on
page 4-48).