Fresenius Medical Care multiFiltrate Instructions For Use Manual page 136

Table of Contents

Advertisement

Chapter 4: Operation
Haemoperfusion (HP)
4.10.3 Inserting the heparin syringe
4.10.4 Complete tubing arrangement
4-94
Note
If a heparin syringe is used, it must be inserted and connected before
connecting the patient.
Note on C
The syringe wings must be placed in the syringe wing slot.
Note on D
The thumbplate of the syringe plunger must be positioned between the
clamps of the grip handle.
Warning
Patient hazard: air embolism
The air detector must be clean and dry.
Do not use any ultrasound-conducting objects or media.
Blood clots (coagula) can cause the air bubble detector to fail.
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
 Insert the heparin syringe according to the
instructions.
 Turn the rotary selector clockwise until the
next screen displays.