Chapter 4: Operation
CRRT treatments with citrate anticoagulation
4.6.9

Disconnecting the patient and removing the tubing system

4-48
Warning
It is forbidden to remove the Ci-Ca lines manually before the patient is
disconnected.
Warning
Consumables must be discarded after the treatment in compliance with
the regulations for the disposal of potentially contaminated materials.
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Indication of the decreasing reinfusion
volume.
Selecting View treatment history? [OK] to
confirm! will display the treatment
parameters.
The optical detector senses non-opaque
fluid.
 When selecting Continue reinfusion, the
remaining reinfusion volume can be
returned to the patient.
or
 Use the rotary selector to select
Terminate reinfusion and press [OK].
The citrate pump stops.
but
The treatment will be terminated
automatically if the reinfusion volume is 0 ml
and the optical detector senses non-opaque
fluid.
The citrate pump stops.