Chapter 4: Operation
Warning
When inserting the tubing systems, observe the following:
The absence of air, kinks, tensions, and twists in the tubing system and
the correct position of all fluid levels. Ensure that supply lines are not
compressed (danger of haemolysis during treatment), since this cannot
always be detected by the safety system. Use the tube holders
provided.
Check the correct position of the screwed connections.
Visually check all solution bags for tightness before connecting them to
the tubing system.
After the blood pump exit, there must not be any clamps placed on the
blood line system.
Warning
Delivery operation of the pump(s) with open doors
(blood pump, substituate pump, filtrate pump, dialysate pump)
When the doors are open and the rotor of the pump(s) is running, make
sure that no objects, such as fingers, hair, or ballpoint pens, come into
contact with the rotor.
Warning
To be observed when working on the tubing system during the
treatment:
If you need to move any part of the tubing system out of position, make
sure the correct layout of the entire tubing system is restored before
continuing treatment, paying special attention to the correct placement
of the positioners.
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
4-3