Fresenius Medical Care multiFiltrate Instructions For Use Manual page 140

Table of Contents

Advertisement

Chapter 4: Operation
Haemoperfusion (HP)
4.10.5.4 Connecting the patient
4-98
Tip
If the patient is not yet available, then recirculation can be continued by
pressing the [START/RESET] key.
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
If using an NaCl solution with two
connectors:
 Disconnect the arterial patient line from
the NaCl solution and connect it to the
arterial vascular access.
or
 Connect the arterial and venous patient
line to the vascular access.
If using an NaCl solution with one
connector:
 Connect the arterial and venous patient
line to the vascular access.
and
 Use the rotary selector to select Start
connection? [OK] to confirm! and press
[OK].
The blood pump will deliver at the
programmed rate.
Settable rate:
Cassette / adults 10 ml/min to
100 ml/min (default 100 ml/min)