Chapter 5: Alarm processing
If the blood leak detector detects blood in the
filtrate line, this is indicated by an audible
signal and a message.
The blood pump is stopped.
 Continue the treatment by pressing the
[START/RESET] key.
The blood pump is started.
The override time for the blood leak
detector is 2 minutes.
The blood leak detector can be deactivated
for MPS during the override time in the
Treatment menu:
 Deactivate the blood leak detector with the
[Blood leak detector] button.
After the safety prompt to deactivate the
blood leak detector has been confirmed with
[Yes], the blood leak detector no longer
monitors for haemolysis or blood loss (Blood
leak detector deactivated!). Monitoring can
be reactivated at any time in the Treatment
menu.