Chapter 4: Operation
CRRT treatments

4.5.10 End of treatment

4.5.10.1 Terminating the treatment

4.5.10.2 Starting reinfusion

4-30
 Press the [ESC] key.
 Use the rotary selector to select End of
treatment from the menu bar and press
[OK] to confirm.
 Stop the treatment by pressing the
[STOP] key.
or
 Use the [ESC] key to select a different
menu from the menu bar.
 Continue the treatment by pressing the
[START/RESET] key.
or
 Disconnect the arterial patient line from
the patient and connect it to the NaCl
solution.
 Use the rotary selector to select Start
disconnection? [OK] to confirm! and
press [OK].
When selecting Disconnect without blood
return, the program will go directly to
"Disconnecting the patient" without reinfusion
after a safety prompt (see chapter 4.5.11 on
page 4-32).
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
The blood pump is running.
Press the [STOP] key for approx.
3 seconds.
The venous clamp closes.
The blood pump is stopped.