4.9.6

Treatment

4.9.6.1
Treatment main screen
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Note
Once treatment has started and the blood pump has achieved its target
delivery rate, the alarm limits are set automatically after approx.
10 seconds. Following this, the pressure alarm limits must be
checked by the operator and adjusted individually as required.
Slow continuous ultrafiltration (SCUF)
The optical detector senses opaque fluid.
The blood pump is stopped.
If not done yet when using an NaCl
solution with two connectors:
 Disconnect the venous patient line from
the NaCl solution and connect it to the
venous vascular access.
 Use the rotary selector to select Start
treatment? [OK] to confirm! and press
[OK].
The main screen is displayed throughout the
entire treatment.
Depending on the treatment mode, the menu
field shows:
The current pressures (arterial, venous,
TMP)
The current flow rates (blood flow,
ultrafiltration)
Heparin
The status bar shows:
The treatment mode
The progression of the treatment time
The balance
Continuous anticoagulation on / off
Chapter 4: Operation