Chapter 4: Operation
Membrane plasma separation (MPS)
4.8.6.6

Connecting the patient

4-72
If using an NaCl solution with two
connectors:
 Disconnect the arterial patient line from
the NaCl solution and connect it to the
arterial vascular access.
or
 Connect the arterial and venous patient
line to the vascular access.
If using an NaCl solution with one
connector:
 Connect the arterial and venous patient
line to the vascular access.
and
 Use the rotary selector to select Start
connection? [OK] to confirm! and press
[OK].
The optical detector senses opaque fluid.
If not done yet when using an NaCl
solution with two connectors:
 Disconnect the venous patient line from
the NaCl solution and connect it to the
venous vascular access.
 Use the rotary selector to select Start
treatment? [OK] to confirm! and press
[OK].
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
The blood pump is stopped.