Chapter 4: Operation
Membrane plasma separation (MPS)
4.8.6.6
Connecting the patient
4-72
If using an NaCl solution with two
connectors:
Disconnect the arterial patient line from
the NaCl solution and connect it to the
arterial vascular access.
or
Connect the arterial and venous patient
line to the vascular access.
If using an NaCl solution with one
connector:
Connect the arterial and venous patient
line to the vascular access.
and
Use the rotary selector to select Start
connection? [OK] to confirm! and press
[OK].
The optical detector senses opaque fluid.
If not done yet when using an NaCl
solution with two connectors:
Disconnect the venous patient line from
the NaCl solution and connect it to the
venous vascular access.
Use the rotary selector to select Start
treatment? [OK] to confirm! and press
[OK].
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
The blood pump is stopped.