4.5.6.5
Connecting the patient
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Tip
If the patient is not yet available, then recirculation can be continued by
pressing the [START/RESET] key.
Note
In the case of CVVH or CVVHDF, connect the substituate line in
predilution, if necessary.
If using an NaCl solution with two
connectors:
Disconnect the arterial patient line from
the NaCl solution and connect it to the
arterial vascular access.
or
Connect the arterial and venous patient
line to the vascular access.
If using an NaCl solution with one
connector:
Connect the arterial and venous patient
line to the vascular access.
and
Use the rotary selector to select Start
connection? [OK] to confirm! and press
[OK].
The blood pump will deliver at the
programmed rate.
Settable rate:
Cassette / adults 10 ml/min to
100 ml/min (default 100 ml/min)
Chapter 4: Operation
CRRT treatments
4-25