Filling Plasma - Fresenius Medical Care multiFiltrate Instructions For Use Manual

Table of Contents

Advertisement

4.8.6.5

Filling plasma

Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Tip
When using Fresenius fresh frozen plasma bags (FFP), for example,
the plasma pole (part no. M28 004 1) can be used if required.
Membrane plasma separation (MPS)
 Stop recirculation by pressing the [STOP]
key.
Press the [STOP] key for approx.
3 seconds.
It is imperative to enter the net volume of the
replacement solution.
 Use the rotary selector to select Start
filling plasma? [OK] to confirm! and
press [OK].
The blood pump will deliver at the
programmed rate.
Settable rate:
10 ml/min to 100 ml/min (default
100 ml/min)
Time-controlled filling of plasma is in
progress.
Chapter 4: Operation
4-71