Chapter 14: Options
Citrate / calcium rates not matching
Balancing off (consequence: calcium pump also stops)
HF bag change (consequence: balancing off, calcium pump also stops)
Syringe change (consequence: balancing off, calcium pump also stops)
Ci-Ca bag change (consequence: balancing off, calcium pump also stops)
If calcium bag change is selected
If citrate bag change is selected
14-2
A warning will be emitted that the rates are not matching and that the
ratios / rates have to be adjusted appropriately. If the warning is
confirmed without changing the parameters, it will be displayed again
every 2 minutes. If the rate is adjusted, but is still not matching, the
warning will be displayed again immediately.
After 5 minutes, a warning will be emitted, informing the operator that
calcium substitution has been interrupted. After another 5 minutes, an
alarm will be emitted, which will be repeated every 2 minutes.
(see Balancing off (consequence: calcium pump also stops) on
page 14-2)
(see Balancing off (consequence: calcium pump also stops) on
page 14-2)
(see Balancing off (consequence: calcium pump also stops) on
page 14-2)
An alarm will be emitted if bag change is not completed within 2 minutes
of selecting the calcium bag change. This alarm will be repeated every
2 minutes.
An alarm will be emitted if bag change is not completed within 2 minutes
of selecting the citrate bag change. This alarm will be repeated every
2 minutes.
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015