Chapter 10: Transport / storage
Locking
Relocation inside buildings during preparation (recirculation)
Before relocating
10-2
The cart comprises 4 wheels. The two wheels on the front (A) are each
equipped with a break. The device can be swiveled, turned, or moved
in any direction by means of the two handles on the upper part (monitor
hood).
All breaks must be locked if the device is in its final position or in
operation.
After the device has been set up, it may only be relocated inside a
building or ward.
In other words:
– The functional test is complete.
– The tubing systems (cassette) have been inserted, primed, and
rinsed.
– The treatment parameters have been entered.
– The device is in the "Recirculation" mode.
– The loads on the device must not exceed the following
specifications. Preferably use the front hooks.
Left IV pole
Substituate / dialysate scale, each
Right IV pole
The device can be disconnected from the power supply by pulling out
the power plug. The device indicates a power failure. The audible alarm
can be suppressed for 2 minutes by pressing the [Mute] key. The
device must be relocated as fast as possible, as battery operation is
only possible for a limited time.
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
A
5 kg
12 kg
6 kg