6.13 RECall Group
The commands in this group deal with outputting stored measured data and loading setup parameters. You can make
the same settings that you can by pressing SAVE/LOAD on the front panel and using the LoAd menu. There are no
front panel keys that output measured data that is stored.
:RECall:NUMber?
Function
Queries the number of blocks of measured data
that is stored.
Syntax
:RECall:NUMber?
Example
:RECALL:NUMBER? -> 600
:RECall[:NORMal]:VALue?
Function
Queries the numeric data at the specified block
number.
Syntax
:RECall[:NORMal]:VALue? {<NRf>}
<NRf> = 1 to 9000 (block number)
Description • Always specify <NRf>. The numeric data at the
specified block number will be returned.
• If you omit <NRf> or specify a number greater
than the number of blocks that contain stored
measured data (the number returned by :
RECall:NUMber?), the entire returned numeric
data will be "NAN" (no data).
• The output items and format are the same as
those of ":NUMeric[:NORMal]:VALue? (when
the item number is not specified)." To set the
output items and format, use the NUMeric
group commands.
:RECall:LIST:VALue?
Function
Queries the numeric list data of harmonic
measurement at the specified block number.
Syntax
:RECall:LIST:VALue? {<NRf>}
<NRf> = 1 to 600 (block number)
Description • This is only valid on models with the harmonic
measurement (/G5) option.
• Always specify <NRf>. The numeric list data at
the specified block number will be returned.
• If you omit <NRf> or specify a number greater
than the number of blocks that contain stored
measured data (the number returned by :
RECall:NUMber?), the entire returned numeric
list data will be "NAN" (no data).
• The output items and format are the same
as those of ":NUMeric:LIST:VALue? (when
the item number is not specified)." To set the
output items and format, use the NUMeric
group commands.
:RECall:PANel
Function
Loads a setup parameter file.
Syntax
:RECall:PANel {<NRf>}
<NRf> = 1 to 4 (file number)
6-30
IM WT310E-17EN