To Increase The Start Frequency; To Set The Sweep Time To Auto Mode; To Widen The System Bandwidth - HP 8712ES User Manual

Rf network analyzers
Hide thumbs Also See for HP 8712ES:
Table of Contents

Advertisement

Table 5-1
Optimizing Measurements
Increasing Sweep Speed

To Increase the Start Frequency

Since the analyzer sweeps frequencies below approximately 20 MHz at a
slower rate, you can increase the start frequency to speed up the sweep.
1. Press
Start
FREQ
2. Enter the highest start frequency possible for your measurement.

To Set the Sweep Time to AUTO Mode

Auto sweep time mode (the factory preset instrument mode) maintains
the fastest sweep speed possible for any particular measurement
settings.
1. Press
SWEEP
label. When AUTO is all capital letters, it indicates that the analyzer
is in auto sweep time mode. If MAN is all capital letters, the analyzer
is in manual sweep time mode.
2. If necessary, press
to AUTO.

To Widen the System Bandwidth

Press
AVG
System Bandwidth
bandwidth increases, the sweep speed gets faster and the trace noise
increases.
Relationship between System Bandwidth and Sweep Speed
System Bandwidth
Wide
Medium Wide
Medium
Medium Narrow
Narrow
Fine
5-4
.
and look at the
Sweep Time AUTO man
Sweep Time auto MAN
to widen the IF bandwidth. As the
fastest
faster
fast
slow
slower
slowest
to toggle the time mode
Sweep Speed