TIP
Table 3-2
Measuring S-Parameters using a
Two-Port Calibration
This section describes how to perform a two-port calibration. Once done,
this calibration can be used with all types of S-parameter measurements.
It should be noted that other types of error correction, though not as
accurate as a two-port calibration, allow faster measurement speed.
Therefore, if measurement speed is of more importance than highest
accuracy, you may want to use a different type of error correction than
the two-port calibration. See the table below for a list of all possible
calibration choices for S-parameter measurements.
An optimum calibration is critical for achieving best measurement
accuracy. Refer to
Chapter 6, "Calibrating for Increased Measurement
Accuracy,"
for detailed information about all of the various aspects of
calibration.
Calibration Choices for S-Parameter Measurements
Measurement Type
S
Refl Port 1 or
11
S
Refl Port 2
22
S
Fwd Trans or
21
S
Rev Trans
12
ES User's Guide
Measuring S-Parameters using a Two-Port Calibration
Calibration Choices
Default 1-Port
Default 2-Port
User 1-Port
User 2-Port
Default Response
Default 2-Port
User Response:
Response
Response & Isolation
Enhanced Response
User 2-Port
Normalize
Making Measurements
3-19