View and Interpret the Power Measurement
Results
1. To view the measurement trace, press
2. To interpret the power measurement, refer to
analyzer's display if you are making this measurement on your
instrument.
a. When making a power measurement, the display shows the power
measured at the analyzer's input connector. This power is absolute
power, as opposed to a power ratio.
b. Note that when making a power measurement, the values
associated with the vertical axis are in units of dBm, which is the
power measured in reference to 1 mW.
0 dBm = 1 mW
• −10 dBm = 100 µW
• +10 dBm = 10 mW
Other units available for a power measurement include dBW,
dBµW, W, mW and µW.