Using Instrument Functions
Using Markers
To Use Delta (∆) Marker Mode
In delta marker mode, a reference marker is placed at the active marker
position. All marker values are then displayed in reference to this delta
marker. When the amplitude of the measurement trace changes, the
reference marker value also changes. The delta marker is represented by
a triangle pointing up (∆) with a delta symbol (∆) below it rather than a
number.
Limit testing can be performed on delta amplitude and delta frequency
using specific markers. See
page 4-32
chapter.
To follow along with this example, set up the instrument for an S
forward transmission measurement of the bandpass filter that was
shipped with your instrument.
1. Press
MARKER
Mkr
> Max
to move marker 1 (the active marker) to the maximum
value on the trace.
2. Press
Prior Menu Prior Menu Marker Functions
Delta Mkr ON
position.
3. Press
Prior Menu Marker Search Min Search Mkr
to place marker 1 to the minimum point on the trace.
4. The difference between the markers' frequency and amplitude value
is shown in the upper-right corner of the analyzer screen.
5. Press
Prior Menu Prior Menu 2:
move marker 2 towards the right-hand side of the display screen.
Note that the marker 2 values are also in reference to the delta
marker.
4-24
"To Use Marker Limit Functions" on
All Off
Marker Search
to place a reference marker at the active marker
Max Search
. Use the front panel knob to
ES User's Guide