13. Determine the offset delay of the calibration short:
• Press
SYSTEM OPTIONS
• Press
Next Screen
"CAL KIT: Port 2," display, find the numeric value for the delay of
the short.
14. Subtract this value from the value calculated in
corresponds to the delay of the adapter.
15. Remove the short from the adapter.
16. Reselect a type-N(m) calibration kit for port 2. This is necessary
because the type-N(m) short has been removed from the adapter.
Press
More Cal Cal Kit
CAL
the
Type-N(m)
.
17. Modify the calibration kit definition of the through-standard by
entering the electrical delay of the adapter. To do so:
• Press
CAL
Delay
. Enter the delay value with the keypad and softkeys.
18. Save (and automatically assign) the modified cal kit buffer as a
user-defined cal kit for port 2. To do so:
• Press
Prior Menu Mod Kit SaveRecall
knob or the
the display's table. (In this example, highlight Cal Kit A.) Press
.
Save
If you want to enter a cal kit description in the display's table,
press
Add Kit Description
analyzer's DIN KEYBOARD connector to enter the description.
(If you don't have a keyboard, you can use the analyzer's
key to select and enter characters from the top of the analyzer's
display. Press
ES User's Guide
Measuring S-Parameters using a Two-Port Calibration
Operating Parameters
until the Cal Kit screen is displayed. In the
, and use the front panel knob or
keys to highlight Port 2 in the display's table. Press
More Cal Cal Kit Modify Type-N(m)
keys to highlight the desired User Kit in
. Use a keyboard connected to the
keys or the front panel knob and the
Enter
when done.)
Making Measurements
Step
11. This
Thru
. Use the front panel
Select Char
3-31