14. Subtract this value from the value calculated in
corresponds to the delay of the adapter.
15. Remove the short from the adapter.
16. Modify the Port 1 calibration kit definition of the through-standard by
entering the electrical delay of the female-to-female adapter. To do so:
•Press
More Cal Cal Kit
CAL
the
Press
Modify 3.5 mm Thru Delay
the keypad and softkeys.
17. Save (and automatically assign) the modified cal kit buffer as a
user-defined cal kit for port 1. To do so:
• Press
Prior Menu Mod Kit SaveRecall
knob or the
the display's table. (In this example, highlight Cal Kit A.) Press
.
Save
If you want to enter a cal kit description in the display's table,
press
Add Kit Description
analyzer's DIN KEYBOARD connector to enter the description.
(If you don't have a keyboard, you can use the analyzer's
key to select and enter characters from the top of the analyzer's
display. Press
18. Modify the Port 2 calibration kit definition of the through-standard by
entering the electrical delay of the female-to-female adapter. To do so:
•Press
More Cal Cal Kit
CAL
the
Press
Modify Type-N(m)
with the keypad and softkeys.
19. Save (and automatically assign) the modified cal kit buffer as a
user-defined cal kit for port 2. To do so:
• Press
Prior Menu Mod Kit SaveRecall
knob or the
the display's table. (In this example, highlight Cal Kit B.) Press
.
Save
ES User's Guide
Measuring S-Parameters using a Two-Port Calibration
and use the front panel knob or
keys to highlight Port 1 in the display's table.
keys to highlight the desired User Kit in
. Use a keyboard connected to the
keys or the front panel knob and the
Enter
when done.)
and use the front panel knob or
keys to highlight Port 2 in the display's table.
Thru
keys to highlight the desired User Kit in
Making Measurements
Step
12. This
. Enter the delay value with
. Use the front panel
Select Char
Delay
. Enter the delay value
. Use the front panel
3-37